User manual HANNA INSTRUMENTS HI 8410

Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... DON'T FORGET : ALWAYS READ THE USER GUIDE BEFORE BUYING !!!

If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Lastmanuals provides you a fast and easy access to the user manual HANNA INSTRUMENTS HI 8410. We hope that this HANNA INSTRUMENTS HI 8410 user guide will be useful to you.

Lastmanuals help download the user guide HANNA INSTRUMENTS HI 8410.


Mode d'emploi HANNA INSTRUMENTS HI 8410
Download
Manual abstract: user guide HANNA INSTRUMENTS HI 8410

Detailed instructions for use are in the User's Guide.

[. . . ] Instruction Manual HI 8410 Dissolved Oxygen Process Controller www. hannainst. com 1 Dear Customer, Thank you for choosing a HANNA instruments® product. Please read this instruction manual carefully before using the instrument. This manual will provide you with all the necessary information for correct use of the instruments, as well as a precise idea of thier versatility in a wide range of applications. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst. com These instruments are in compliance with the directives. TABLE OF CONTENTS Preliminary Examination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . [. . . ] In order to avoid malfunction leave unconnected the unused terminal. 10 · Alarm Contacts (terminals 5, 6) During normal operation these terminals remain closed. level is not within the tolerance of the set value, the alarm contact is open. See also page 15. · Overtime dosing When enabled this feature ensures that overdosage is avoided. If the dosage relay is active more than the selected period an alarm condition is activated and the dosage relay is deactivated. To leave the overtime dosage alarm status change the position of the OFF/AUTO/ON switch to OFF and then to AUTO again. To set the overtime dosing period rotate the overtime knob to the proper position. 5 min to 60 min (from MIN to MAX position). To disable the overtime dosage feature make a short with a jumper wire between the terminals of the proper connectors. (see picture, terminals 1, 2) · Hysteresis knob Put the hysteresis knob in the desired position. The dosage will be active according to the DO reading, set point value and the hysteresis set value. DOSING MODE SELECTION The DOSAGE switch and the corresponding right side LED on the front panel are used to select and indicate the dosing mode. OFF mode The dosage is disabled. The corresponding DOSAGE MODE LED blinks. 11 AUTO mode The dosage is activated and deactivated according with the selected setpoint. The corresponding DOSAGE MODE LED (right side of the switch) is off. Be sure that the DOSAGE switch is in AUTO position when the meter is in normal operating mode. ON mode The dosage is always activated. The corresponding DOSAGE MODE LED blinks. OPERATIONAL GUIDE INITIAL PREPARATION & INSTALLATION Material needed: · a 3-wire power cable to connect HI 8410 Ensure the controller has been calibrated. probe is immersed in the sample to be tested. reading can be displayed in % air saturation or in ppm (mg/L) according to the range selection switch. For accurate dissolved oxygen measurements a water flow of at least 5-7 cm/sec is required. In this way a constant replenishment of the oxygen-depleted membrane surface is ensured. Accurate readings are not possible in still water. Note: For some particular applications, such as fish farming, the membrane can be sterilized with stabilized iodine (20 to 50 ppm), typically used for this purpose. Note: For use in harsh environments, it is recommended to protect the membrane with the optional HI 76410-0 sleeve. The response time will slightly slow down. 12 Ensure that the HI 76410 DO probe is connected to the meter according with the colors indicated on the mask. OPERATING INFORMATION All parameters are set through the front panel keys and trimmers. [. . . ] When shipping any instrument, make sure it is properly packaged for complete protection. Recommendations for Users Before using these products, make sure that they are entirely suitable for the environment in which they are used. Operation of these instruments in residential area could cause unacceptable interferences to radio and TV equipments, requiring the operator to take all necessary steps to correct interferences. The trimmers are sensitive to electrostatic discharges. It is recommended to use anti-static screwdrivers. [. . . ]

DISCLAIMER TO DOWNLOAD THE USER GUIDE HANNA INSTRUMENTS HI 8410

Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets...
In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service.

Click on "Download the user manual" at the end of this Contract if you accept its terms, the downloading of the manual HANNA INSTRUMENTS HI 8410 will begin.

Search for a user manual

 

Copyright © 2015 - LastManuals - All Rights Reserved.
Designated trademarks and brands are the property of their respective owners.

flag