User manual HANNA INSTRUMENTS HI 8427

Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... DON'T FORGET : ALWAYS READ THE USER GUIDE BEFORE BUYING !!!

If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Lastmanuals provides you a fast and easy access to the user manual HANNA INSTRUMENTS HI 8427. We hope that this HANNA INSTRUMENTS HI 8427 user guide will be useful to you.

Lastmanuals help download the user guide HANNA INSTRUMENTS HI 8427.


Mode d'emploi HANNA INSTRUMENTS HI 8427
Download
Manual abstract: user guide HANNA INSTRUMENTS HI 8427

Detailed instructions for use are in the User's Guide.

[. . . ] 6 mV (ORP) Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 High Impedance Calibration . . . . . . . . . . . . . . . . . . . . 10 pH Values at Various Temperature . . . . . . . . . . 11 Battery Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . [. . . ] Make sure that the LOW BATTERY LED is not lit. If it is lit, change the battery of the simulator before performing the calibration. Verify that the pH meter displays the pH 7 message. If not, adjust the pH meter until the correct reading pH 7 is displayed (refer to the pH meter instruction manual). Select any other pH value on the simulator and set the °C knob to the desired working temperature (e. g. pH 4 at 30°C). pH7 HI 931001: · Connect the simulator to the pH meter using the HI 7858/1 cable provided. · Manually set the temperature compensation of your meter to 25°C. · Switch the simulator on and select the pH mode by pressing the RANGE key. pH pH HI 8424 microcomputer pHmeter ON OFF HI 931001 pH/mV Calibrator SETTING ON OFF RANGE pH HI 8427 pH/mV CALIBRATOR · 0 10 20 30 Make sure that the two additional decimal points are not displayed. If they are displayed, change the battery of the simulator before performing the calibration. Turn the Coarse and Fine knob to set any desired pH value and make sure that the pH meter you are checking reflects the correct result (e. g. 7. 01 pH). · 50 40 °C · · If your pH meter has a temperature probe connected to it, set the °C knob to the measured temperature displayed by your pH meter. If your meter is manually compensated for temperature, adjust the temperature settings for both the pH meter and the simulator to the same temperature. Verify that the pH meter displays the correct pH value compensated at the 6 · If not, adjust the pH meter until the correct reading is displayed (refer to the pH meter instruction manual). pH pH HI 8424 microcomputer pHmeter ON OFF HI 931001 pH/mV Calibrator SETTING 7 mV (ORP) CALIBRATION · HI 8427: · Connect the simulator to the mV (ORP) meter using the HI 7858/1 cable provided. · · pH HI 8427 pH/mV CALIBRATOR Switch the mV meter on and select the desired mV value on the simulator. Make sure that the LOW BATTERY LED is not lit. If it is lit, change the bat-350 tery of the simulator bemV fore performing the calibration. · If the meter is not displaying the correct mV value, adjust the meter until it displays the correct mV (refer to the mV meter instructions manual). Note: The °C knob is not effective during mV calibration. Make sure that the the two additional decimal points are not displayed. If they are displayed, change the battery of the simulator before performing the calibration. Turn the Coarse or Fine knob to any desired value and make sure that the mV meter you are checking reflects the correct result. · If not, adjust the mV meter until the correct reading is displayed (refer to the mV meter instruction manual). mV mV HI 8424 microcomputer pHmeter ON OFF HI 931001 pH/mV Calibrator SETTING Note: Temperature compensation does not apply here. HI 931001: · Connect the simulator to the mV (ORP) meter using the HI 7858/1 cable provided. pH pH HI 8424 microcomputer pHmeter ON OFF HI 931001 pH/mV Calibrator SETTING ON · Switch the simulator on and select the mV mode by pressing the RANGE key. OFF RANGE 8 9 CABLE HIGH IMPEDANCE TEST pH VALUES AT VARIOUS TEMPERATURE HI 8427 only: Temperature has an effect on pH. [. . . ] If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization number from the Customer Service department and then send it with shipping costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection. [. . . ]

DISCLAIMER TO DOWNLOAD THE USER GUIDE HANNA INSTRUMENTS HI 8427

Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets...
In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service.

Click on "Download the user manual" at the end of this Contract if you accept its terms, the downloading of the manual HANNA INSTRUMENTS HI 8427 will begin.

Search for a user manual

 

Copyright © 2015 - LastManuals - All Rights Reserved.
Designated trademarks and brands are the property of their respective owners.

flag