User manual HANNA INSTRUMENTS HI 9033

Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... DON'T FORGET : ALWAYS READ THE USER GUIDE BEFORE BUYING !!!

If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Lastmanuals provides you a fast and easy access to the user manual HANNA INSTRUMENTS HI 9033. We hope that this HANNA INSTRUMENTS HI 9033 user guide will be useful to you.

Lastmanuals help download the user guide HANNA INSTRUMENTS HI 9033.


Mode d'emploi HANNA INSTRUMENTS HI 9033
Download
Manual abstract: user guide HANNA INSTRUMENTS HI 9033

Detailed instructions for use are in the User's Guide.

[. . . ] 4 FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 9034 . . . . . . . . . . 5 FUNCTIONAL DESCRIPTION OF HI 76302W CONDUCTIVITY PROBE. . . . . 6 OPERATIONAL GUIDE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 CALIBRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . [. . . ] · Connect the probe to the meter securely by aligning the pins with the socket, pushing the plug in and tightening the threaded ring. · Immerse the probe into the beaker, while paying attention that the holes on the PVC sleeve are completely submerged. · Turn the instrument on by pressing the ON/OFF key and select the appropriate range (e. g. 19. 99 mS/cm). Note: Probe body and sleeve are made of PVC, and are susceptible to damage due to temperatures exceeding 60°C (140°F). If the probe is exposed to high temperature, the bond between the rings and the probe body may become impaired and the probe will not function properly. In this case, the probe will need to be replaced. ON/OFF 19. 99mS/cm · Tap the probe repeatedly on the bottom of the beaker and stir it to ensure that no air bubbles remain trapped inside the sleeve. · If the probe temperature is close to the solution temperature, the display will stabilize quickly and show the temperature compensated conductivity measurement. However, the ATC circuitry needs a few minutes for temperature compensation if the temperature difference is about 5°C (9°F). 8 9 · When the reading is stable, adjust the calibration trimmer to read the solution conductivity value @25°C. For example if using HI 7030, turn the trimmer to read 12. 88 mS/cm @ 25°C. mS The calibration is now complete and the instrument is ready for use. All subsequent measurements will be compensated to 25°C (77°F). Note: Calibration can also be standardized to have readings compensated to 20°C (68°F), by simply adjusting the trimmer to the calibration solution value at 20°C (68°F). For example, for HI 7030 adjust the trimmer to read 11. 67 mS/cm (see "Conductivity vs. If the instrument cannot be calibrated, refer to the "Probe Maintenance" section. · Tap the probe repeatedly on the bottom of the beaker and stir it to ensure that no air bubbles remain trapped inside the sleeve. · If the probe temperature is close to the solution temperature, the display will stabilize quickly and show the temperature compensated conductivity measurement. However, the ATC circuitry needs a few minutes for temperature compensation if the temperature difference is about 5°C (9°F). · When the reading is stable, adjust the calibration trimmer to read the solution conductivity value @25°C. For example if using HI 7032, turn the trimmer until the display shows 1382 mg/L @25°C. The calibration is now complete and the instrument is ready for use. All subsequent measurements will be compensated to 25°C (77°F). Note: Calibration can also be standardized to have readings compensated to 20°C (68°F), by simply adjusting the trimmer to the calibration solution value at 20°C (68°F). For example, for HI 7032 adjust the trimmer to read 1251 mg/L (see "TDS vs. If the instrument cannot be calibrated, refer to the "Probe Maintenance" section. PROCEDURE FOR HI 9034 · Fill a beaker with 8 cm (3¼") of TDS calibration solution (if possible fill two beakers, one for rinsing the probe and one for calibration). · Connect the probe to the meter securely by aligning the pins with the socket, pushing the plug in and tightening the threaded ring. [. . . ] (03) 3258. 9565 · Fax (03) 3258. 9567 Korea: Tel. (02) 2278. 5147 · Fax (02) 2264. 1729 Malaysia: Tel. (603) 5638. 9940 · Fax (603) 5638. 9829 Singapore: Tel. 6296. 7118 · Fax 6291. 6906 South Africa: Tel. [. . . ]

DISCLAIMER TO DOWNLOAD THE USER GUIDE HANNA INSTRUMENTS HI 9033

Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets...
In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service.

Click on "Download the user manual" at the end of this Contract if you accept its terms, the downloading of the manual HANNA INSTRUMENTS HI 9033 will begin.

Search for a user manual

 

Copyright © 2015 - LastManuals - All Rights Reserved.
Designated trademarks and brands are the property of their respective owners.

flag