User manual ZEISS AXIOLAB POL

Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... DON'T FORGET : ALWAYS READ THE USER GUIDE BEFORE BUYING !!!

If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Lastmanuals provides you a fast and easy access to the user manual ZEISS AXIOLAB POL. We hope that this ZEISS AXIOLAB POL user guide will be useful to you.

Lastmanuals help download the user guide ZEISS AXIOLAB POL.


Mode d'emploi ZEISS AXIOLAB POL
Download
Manual abstract: user guide ZEISS AXIOLAB POL

Detailed instructions for use are in the User's Guide.

[. . . ] Axiolab Pol Microscope Operating manual Unless expressly authorised, forwarding and duplication of this document, utilisa tion and communication of its contents are not permitted. Violations will entail an obligation to pay compensation. All rights reserved in the event of granting of patents or registration of a utility model. This manual is not covered by an update service. Issued by: Carl Zeiss Jena GmbH Zeiss Gruppe Unternehmensbereich Mikroskopie D 07740 Jena Phone: (03641) 64 2936 Fax: (03641) 64 3144 Number of this Operating manual: B 40 016 e Date of issue: 09/95 Axiolab Pol TABLE OF CONTENTS Page Title page Copyright Table of contents . I IV VI VII IX 1 1. 1 1. 2 1. 2. 1 1. 2. 2 1. 3 1. 4 Description Designation, purpose . [. . . ] Never view directly into these high power lamps!The other regulations for your safety, as detailed in the instruction manuals, must be observed strictly!· Adjust the lamp helix or light source image as described in the above mentioned instructions. · Fit the lamp on the stand as described in Section 2. 1 (4). In doing so, orient the lamp housing perpendicular to the stand so that the lamp mount points downwards (no orientation point available). NOTE When using these lamps, the uniformity of illumination for the lens 2. 5x/0. 075 can be improved by slightly readjusting on the focusing screw (symbol with the horizontal double arrow). Check this with a homogeneously reflecting object (mirror), while observing the illumination and the edge of the field. 2 32 09/95 B 40 016 e Axiolab Pol 2. 7. 3 Replacing condensers 1 6 2 5 4 3 Figure 2 25 Replacing the condenser · Raise the rotary stage Pol to the stop by means of the coarse drive (2 25/2). · Lower the condenser carrier by means of the condenser drive (2 25/1) and undo the clamping screw (2 25/3). IMPORTANT When using the overview unit (445313) or the circular polariser (453623), make sure that the rotary bearings of these units do not press on the luminous field diaphragm (2 25/4). · Pull the condenser towards you out of the guide fork; if necessary, tilt down the front optics (shift lever (2 25/5 or 6)). · Insert the other condenser in the reverse order. B 40 016 e 09/95 2 33 Axiolab Pol 2. 7. 4 Fitting and replacing the polariser 1 2 12 13 3 11 10 4 9 14 5 7 8 6 Figure 2 26 Polariser assembly · The swivelling polariser (2 26/5) is secured at two screw on points (2 26/2 and 3) under the condenser carrier (2 26/1). · For installation or removal, use the ball head screwdriver (2 26/6) and a normal screwdriver. For installation, first secure the screw (2 26/4) at the screw in point (2 26/3) and then the screw (2 26/7) at the screw in point (2 26/2). · First secure the polariser D, rotary (453620) (2 26/14) at the screw in point (2 26/11) using the key (2 26/8) and then secure the latching rod (2 26/10) at the screw in point (2 26/12). · Carry out disassembly in the reverse order. NOTE The polarisers have been optimally aligned at the works. In certain circumstances, slight readjustment analogous to Section 2. 7. 6 may be necessary during the course of replacement. 2 34 09/95 B 40 016 e Axiolab Pol 2. 7. 5 Assembling and using the overview unit, swivelling for lens 1. 25x (accessory 445313 for equipment 0. 4 and 0. 5) For overview observation in a specimen field Ø = 16 mm (when using the Plan Neofluar 1. 25x/0. 04 lens / accessory 440300) in transmitted light bright field or polarisation mode, the above mentioned overview unit is needed to illuminate the field. · To facilitate assembly, detach the rotary stage from the stand as described in Section 2. 7. 1 (3). · Using a screwdriver, first unscrew the fillis ter head screw (2 27/2) on the underside of the polariser D, rotating (2 27/1). · Screw the analogous bearing screw of the large field lens (2 27/4) into the now exposed threaded hole in the rotary bearing of the polariser (use key (2 27/5) to do this). NOTE Re insert the spring washer (2 27/3) on assembly. 1 6 2 3 4 8 5 Figure 2 27 Using the overview unit 7 · Screw on the extended latching rod (2 27/6) analogously to the polariser D, rotating (Section 2. 7. 4) after removing the latching rod (2 27/7). · Insert the condenser as described in Section 2. 7. 3 and raise it by means of the condenser drive (2 25/1). · Position the rotary stage on the stand as described in Section 2. 7. 1 (3) and slide it up until the bottom stop (2 19/3) takes effect, and secure it here. [. . . ] W. : Microscopic idenfication of minerals MC Graw Hill, New York 1965 Kordes, E. : Optische Daten zur Bestimmung anorganischer Substanzen mit dem Polarisationsmikroskop Verlag Chemie, Weinheim 1960 Leeder, O. und Blankenburg, H J. : Polarisationsmikroskopie Deutscher Verlag für Grundstoffindustrie, Leipzig 1989 Mosebach, R. : Das Messen optischer Gangunterschiede mit Drehkompensatoren Heidelberger Beitr. und Tertsch, H. : Geometrische Kristallografie und Kristalloptik und deren Arbeitsmethoden Springer Verlag, 3. Wien 1958 Ramdohr, P. : Die Erzminerale und ihre Verwachsungen, 3. [. . . ]

DISCLAIMER TO DOWNLOAD THE USER GUIDE ZEISS AXIOLAB POL

Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets...
In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service.

Click on "Download the user manual" at the end of this Contract if you accept its terms, the downloading of the manual ZEISS AXIOLAB POL will begin.

Search for a user manual

 

Copyright © 2015 - LastManuals - All Rights Reserved.
Designated trademarks and brands are the property of their respective owners.

flag