If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Lastmanuals provides you a fast and easy access to the user manual EUTECH INSTRUMENTS 2-WIRE PHORP TRANSMITTER. We hope that this EUTECH INSTRUMENTS 2-WIRE PHORP TRANSMITTER user guide will be useful to you.
Lastmanuals help download the user guide EUTECH INSTRUMENTS 2-WIRE PHORP TRANSMITTER.
Manual abstract: user guide EUTECH INSTRUMENTS 2-WIRE PHORP TRANSMITTER
Detailed instructions for use are in the User's Guide.
[. . . ] EUTECH INSTRUMENTS
4
Installation
2-Wire pH or ORP Transmitter
Operating Instructions
Operating with one or several Transmitters:
Figure 1 - Operating with 1 Transmitter
1
Preface
Thank you for purchasing the Eutech pH or ORP Transmitter. [. . . ] Blk 55 Ayer Rajah Crescent #04-14/24, Singapore 139 949. Tel: (65) 778 6876; Fax: (65) 773 0836; e-mail: marketing@eutechinst. com; Home page: http://www. eutechinst. com
Figure 2 - Operating with several transmitters
5
Electrical Connections
2
Introduction
The Transmitter is used for the continuous measurement of pH, ORP (REDOX), Conductivity, Dissolved Oxygen or Temperature, in conjunction with an appropriate sensor. Data output is via 4-20 mA current output. The Transmitter is housed in a IP65 enclosure, with openings for input and 4-20mA output. NOTE: In electrically "noisy" environments, select `Symmetrical' mode of operation and connect a stainless steel rod (Potential Matching Pin - PMP) as shown above. Ensure both the PMP and pH or ORP electrodes are immersed into the solution, even when performing calibration.
3
Diagram and Dimensions (in mm)
6
Before Calibration
Remove cover of Transmitter and ensure that electrode is correctly connected as above. Connect Amp meter to the 4/20mA output of Transmitter (Appendix 1).
7
1) 2)
pH Transmitter Calibration
Calibrate in pH 7. 00 buffer first, and then in either pH 4 or pH 10. First, immerse clean pH electrode into a container with fresh pH 7. 0 buffer solution;
68X256901 Rev 1 10/99
3) 4) 5) 6) 7) 8) 9)
Amp meter should read 12mA; if not, calibration is necessary; Locate "Offset adjustment" trimmer; Using a fine screwdriver, slowly turn trimmer till Amp meter reads 12mA; For greater accuracy, adjust slope with pH 4 or 10; Rinse probe in de-ionised water and immerse in either pH 4. 01 or pH 10. 01 buffer solution; In pH 4. 01, the output is 8. 54mA; In pH 10. 01, the output is 15. 44mA; Using a fine screwdriver, slowly turn trimmer, marked `slope', till Amp meter reads correct mA value with respective buffer solutions.
Figure 4 - Method Two
11
pH
Ordering Information
Item Cat Code ECPHXTR ECARTSO05B ECARGTSO05B
10) Remember to disconnect Amp meter and close cover. [. . . ] for 24V
Item 120V Power Supply 220V Power Supply Cat Code EC-120-XTR EC-220-XTR
Housing
77 x 77 x 28 mm, field mountable
10
Appendix 1
Connecting Amp Meter for Calibration
Figure 3 - Method One
68X256901 Rev 1 10/99 [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE EUTECH INSTRUMENTS 2-WIRE PHORP TRANSMITTER
Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service.