If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Lastmanuals provides you a fast and easy access to the user manual EUTECH INSTRUMENTS ALPHA CON 1000 CONDUCTIVITY CONTROLLERTRANSMITTER. We hope that this EUTECH INSTRUMENTS ALPHA CON 1000 CONDUCTIVITY CONTROLLERTRANSMITTER user guide will be useful to you.
Lastmanuals help download the user guide EUTECH INSTRUMENTS ALPHA CON 1000 CONDUCTIVITY CONTROLLERTRANSMITTER.
Manual abstract: user guide EUTECH INSTRUMENTS ALPHA CON 1000 CONDUCTIVITY CONTROLLERTRANSMITTER
Detailed instructions for use are in the User's Guide.
[. . . ] Instruction Manual
lpha-CON1000 Conductivity Controller/Transmitter
Conductivity Controller
lpha CON1000
MEAS
8. 08
25. 0
CAL
REL A REL B ENTER AUTO MANU
mS
o
C
ATC
ALARM
REL A
REL B
ESC
68X216802
This manual serves to explain the use of the lpha-CON1000 series Conductivity controller/transmitter. The manual functions in two ways: firstly as a step by step guide to help the user operate the instrument. Secondly, it serves as a handy reference guide. This instruction manual is written to cover as many anticipated applications of the lpha-CON1000 Conductivity controller/transmitter. [. . . ] 6) Continue with additional Advanced Set-up procedures, or return to the Measurement mode by pressing the and keys (escape) simultaneously.
HOLD SETUP
on
Atc
HOLD SETUP
0. 0
25. 0
o
C
ATC
7. 2. 1
Setting manual temperature compensation Note: This parameter is blanked out when the controller is set for HOLD SETUP ATC operation. For manual temperature compensation, you can set two different temperatures: process and calibration. This allows calibration at a Atc temperature other than your process temperature. Example: setting a calibration temperature of 25 oC lets you calibrate using standard buffer solutions at 25 oC, even if your process temperature is a HOLD SETUP different temperature. 1) Select "ATC oFF" as described above in section 7. 4. 1 2) Press the ENTER key. The upper display shows the current P. C process temperature and the lower display shows "P. oC" to indicate process temperature.
oFF
25. 0
HOLD SETUP
25. 0
C. C
5) 6) 7)
3) Press the or keys to adjust the process temperature value. You can adjust the value from 9. 9 to 125 oC. 4) After you set the process temperature value, press the ENTER key. The upper display shows the current calibration temperature and the lower display shows "C. oC" to indicate calibration temperature.
Press the or keys to adjust the calibration temperature value. You can adjust this value from 9. 9 to 125 oC. Press the ENTER key to confirm your selection. Continue with additional Advanced Set-up procedures, or return to Measurement mode by pressing the and keys (escape) simultaneously.
68X216802
Operating Instructions
lpha-CON1000
7. 3
Control Relay A/Control Relay B (SP1/SP2) sub-function
The SP1 option sets the operating parameters for Relay A; and SP2 for relay B. Since these groups have the same set-up parameters, they are described together.
HOLD SETUP HOLD SETUP HOLD SETUP HOLD SETUP HOLD SETUP HOLD SETUP
SP 1
7. 3. 1
SP 1
100
SP 1
Lo
HYS
20
On. d
0
Of. d
0
Entering the Set point 1 (Set point 2) sub-function 1) Enter Advanced Set-up mode. Push the ENTER key and HOLD SETUP scroll to Advanced Set-up security code "22". Push the ENTER key again.
SP 1
2) Press the or keys to scroll until the upper display shows SP1 (SP2).
HOLD SETUP
100
SP 1
HOLD SETUP
µS
Lo
SP 1
7. 3. 2 Selecting the set point values This lets you choose the value that will cause your controller to activate (Default: SP1 = 6. 20mg/l; SP2 = 4. 00mg/l). 1) Follow directions in 7. 5. 1 to enter Control Relay mode. If you are in this mode, skip to step 2. The upper display shows the current set point value and the lower display shows SP1 (SP2). 3) Press the or keys to select your value for Set point 1 (Set point 2). Your controller will activate at the value you select. 4) Press the ENTER key to confirm your selection. 5) Proceed to 7. 5. 3, or return to Measurement mode by pressing the and keys simultaneously (escape).
7. 3. 3 Choosing High or Low set points Select a low set point to activate controller when the dissolved oxygen value undershoots the low set point; select a high set point to activate controller when the value overshoots the high set point. [. . . ] 0 to 10% of Full Scale 2 potential-free change-over contacts max. 600 VA 110 / 220 VAC (jumper selectable) 48 to 62 Hz 0/4 to 20 mA, galvanically isolated max. 600 Terminal blocks 5-pole / 17-pole, removable slow-blow 250 V / 100 mA According to EN 50081-1 According to EN 50082-1 0 to 50 oC 10 to 95%, non-condensing 175 x 96 x 96 mm max. 0. 7 kg ABS with polycarbonate (front housing) IP 54 / IP 65
68X216802
25
10
Accessories
Assembly Accessories
Product Description Code no. [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE EUTECH INSTRUMENTS ALPHA CON 1000 CONDUCTIVITY CONTROLLERTRANSMITTER
Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service.