If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Lastmanuals provides you a fast and easy access to the user manual EUTECH INSTRUMENTS ALPHA DO 1000 DISSOLVED OXYGEN CONTROLLERTRANSMITTER. We hope that this EUTECH INSTRUMENTS ALPHA DO 1000 DISSOLVED OXYGEN CONTROLLERTRANSMITTER user guide will be useful to you.
Lastmanuals help download the user guide EUTECH INSTRUMENTS ALPHA DO 1000 DISSOLVED OXYGEN CONTROLLERTRANSMITTER.
Manual abstract: user guide EUTECH INSTRUMENTS ALPHA DO 1000 DISSOLVED OXYGEN CONTROLLERTRANSMITTER
Detailed instructions for use are in the User's Guide.
[. . . ] Instruction Manual
lpha-DO 1000
Dissolved Oxygen Controller/Transmitter
Technology Made Easy . . .
68X242101 Rev 3, 06/04
Preface
This manual serves to explain the use of the lpha-DO1000 series Dissolved Oxygen controller/transmitter. The manual functions in two ways: firstly as a stepby-step guide to help the user operate the instrument. Secondly, it serves as a handy reference guide. This instruction manual is written to cover as many anticipated applications of the lpha-DO1000 Dissolved Oxygen controller/transmitter. [. . . ] A Compare the current measured temperature on the controller display to a thermometer known to be accurate. Note down the correct temperature value. HOLD SETUP °C Press the or keys to scroll the lower display to match the correct value. The upper display will now show the offset value. You can offset the temperature 25. 0 values to ± 5 °C. Press the ENTER key to confirm your selection. Continue with additional Advanced Set-up procedures, or return to the Measurement mode by pressing the or keys (escape) simultaneously.
HOLD SETUP
se
HOLD SETUP
°c
2. 6.
17
Operating Manual
alpha DO 1000
7. 4. 3
Setting Manual Temperature Compensation Note: This parameter is blanked out when the controller is set for ATC operation.
For manual temperature compensation, you can set two different temperatures: process and calibration. This allows calibration at a temperature other than your process temperature. Example: setting a calibration temperature of 25 °C lets you calibrate using standard buffer solutions at 25 °C, even if your process temperature is a different temperature. Select "ATC oFF" as described above in section 7. 4. 1 2. The upper display shows the current process temperature and the lower display shows "P. °C" to indicate process temperature. Press the or keys to adjust the process temperature value. You can adjust the value from HOLD SETUP 9. 9 to 125 °C. After you set the process temperature value, press the ENTER key. c calibration temperature and the lower display shows "C. °C" to indicate calibration temperature. Press the or keys to adjust the calibration temperature value. You can adjust this value from °C 9. 9 to 125 °C. Press the ENTER key to confirm your selection. Continue with additional Advanced Set-up procedures, or return to Measurement mode by pressing the or keys (escape) simultaneously.
FF
25. 0 25. 0
7. 5
Control Relay A/Control Relay B (SP1/SP2) sub-function
The SP1 option sets the operating parameters for Relay A; and SP2 for relay B. Since these groups have the same set-up parameters, they are described together.
HOLD SETUP HOLD SETUP HOLD
mg/l
SETUP
HOLD
SETUP
HOLD
mg/l
SETUP
HOLD
SETUP
Sp1
6. 20
Sp1
L
Sp1
02 . 0
Hys
O.
0
OF.
0
18
Operating Manual
alpha DO 1000
7. 5. 1 1.
Entering the Set point 1 (or Set point 2) sub-function Enter Advanced Set-up mode. Push the ENTER key and (use or keys to) scroll to Advanced Set-up security code "22". [. . . ] of calibration points Flow Rate Response Time Temperature Resolution Relative Accuracy Sensor Temperature Response Temperature Compensation Pressure input Salinity input range Set-point and Controller Functions Controller characteristics Adjustable period with pulse length controller Adjustable period with pulse frequency controller Pickup / Dropout delay Switching D. O. hysteresis Contact outputs, controller Switching voltage Switching current Switching power Alarm Functions Function (selectable) Pickup delay Switching voltage Switching current Switching power Electrical Data and Connections Power Requirements Frequency Signal Output Load D. O. input Input impedance Cable length Connection terminal Mains fuse / fine wire fuse
32
Operating Manual EMC Specifications Emissions Susceptibility Environmental Conditions Ambient temp. operating range Relative humidity Mechanical Specifications Dimensions (control panel housing - L x H x W) Weights (control panel housing) Material Insulation (Front / Housing)
alpha DO 1000
According to EN 50081-1 According to EN 50082-1 0 to 50 °C 10 to 95%, non-condensing 175 x 96 x 96 mm max. [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE EUTECH INSTRUMENTS ALPHA DO 1000 DISSOLVED OXYGEN CONTROLLERTRANSMITTER
Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service.