Detailed instructions for use are in the User's Guide.
[. . . ] Alpha COND 560
Controller mS / µS / ºC / ºF
ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. patent 6, 793, 787.
AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perpHect, PerpHecT, PerpHecTion, pHISA, pHuture, Pure Water, Sage, Sensing the Future, SensorLink, ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS and TURBO2 are registered trademarks of Thermo Fisher.
1-888-pHAX-ION, A+, All in One, Aplus, AQUAsnap, AssuredAccuracy, AUTOBAR, AUTO-CAL, AUTO DISPENSER, Auto-ID, AUTO-LOG, AUTO-READ, AUTOSTIR, Auto-Test, BOD AutoEZ, Cable-Free, CERTI-CAL, CISA, DataCOLLECT, DataPLUS, digital LogR, DirectCal, DuraProbe, Environmental Product Authority, Extra Easy/Extra Value, FAST QC, GAP, GLPcal, GLPcheck, GLPdoc, ISEasy, KAP, LabConnect, LogR, Low Maintenance Triode, Minimum Stir Requirement, MSR, NISS, One-Touch, One-Touch Calibration, One-Touch Measurement, Optimum Results, Orion Star, Pentrode, pHuture MMS, pHuture Pentrode, pHuture Quatrode, pHuture Triode, Quatrode, QuiKcheK, rf link, ROSS Resolution, SAOB, SMART AVERAGING, Smart CheK, SMART STABILITY, Stacked, Star Navigator 21, Stat Face, The Enhanced Lab, ThermaSense, Triode, TRIUMpH, Unbreakable pH, Universal Access are trademarks of Thermo Fisher.
Guaranteed Success and The Technical Edge are service marks of Thermo Fisher.
PerpHecT meters are protected by U. S. PerpHecT ROSS electrodes are protected by U. S. ORION Series A meters and 900A printer are protected by U. S. [. . . ] Press or key to select the desired value. Select temperature coefficient for calibration liquid: The lower display shows `C. tC'. The upper display shows last configured temperature coefficient for calibration liquid (if any). Press or key to select the desired value. The controller reverts to tC screen.
Press or key to access other setup screens or press and key simultaneously (escape) to return to measurement mode.
5. 3
Temperature Settings
Automatic Temperature Compensation (ATC): The conductivity value of a sample is affected by temperature. Use ATC feature of the controller to compensate for conductivity changes when the temperature of the sample or process liquid fluctuates. Connect a separate temperature probe (Pt100/pt1000) to the controller if an integrated conductivity/Temperature probe is not available. Manual Temperature Compensation (MTC): Set the controller to MTC (disable ATC) when the temperature of sample or process liquid is constant and a temperature probe is not available. If you disable ATC, the controller allows you to set your process temperature (P. ºC) and calibration temperature (C. ºC). This allows calibration at a different temperature other than the process temperature. Example: Setting calibration temperature of 25°C lets you calibrate using standard solutions at 25°C, even if your process temperature is different from 25°C.
4
SETUP SETUP
1
SETUP
HOLD
2
ENT
SETUP
HOLD °C
3
SETUP
5
ENT
HOLD
ENT
HOLD °C
ENT
HOLD °C ATC °C ATC
2
SETUP
HOLD °F
3
SETUP
HOLD
6
ENT
SETUP
HOLD
7
SETUP
ENT
HOLD
ENT
ENT
°C
1
From measurement mode press ENT key to enter setup mode as described in section 5. 1. The LCD shows the first screen of setup mode (rAng). Press or key to select Temperature settings screen (SET ºCF). Press ENT key to access temperature settings (SET ºCF). Selecting unit of measurement for temperature: The upper display shows `Unit' and the lower display shows the last configured unit of measurement for temperature. Press or key to select the desired units for temperature (ºC or ºF). Press ENT key to confirm your selection.
2
3
Enable/disable ATC: The lower display shows `AtC' and the upper display shows the last configured ATC selection (`On' or `OFF').
Press or key to enable (ATC On) or disable (ATC OFF) automatic temperature compensation. Press ENT key to confirm your selection. If ATC enabled (ATC On):
4
Selecting temperature offset: The upper display shows the last configured temperature offset value (if any), otherwise the default is zero. The lower display shows currently measured temperature reading (including last configured offset value). LCD shows `ATC' annunciator in lower-right corner. Place a thermometer, which is known to be accurate, in your sample or process liquid. [. . . ] The warranty card must be endorsed by the Authorized Distributor at the point of sale. If repair or adjustment is necessary and has not been the result of abuse or misuse within the designated period, please return freight pre-paid and correction will be made without charge. Thermo Scientific/ Eutech will determine if the product problem is due to deviations or customer misuse. Out of warranty products will be repaired on a charged basis. [. . . ]