Detailed instructions for use are in the User's Guide.
[. . . ] Alpha pH 550
Monitor pH / ORP
ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. patent 6, 793, 787.
AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perpHect, PerpHecT, PerpHecTion, pHISA, pHuture, Pure Water, Sage, Sensing the Future, SensorLink, ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS and TURBO2 are registered trademarks of Thermo Fisher.
1-888-pHAX-ION, A+, All in One, Aplus, AQUAsnap, AssuredAccuracy, AUTOBAR, AUTO-CAL, AUTO DISPENSER, Auto-ID, AUTO-LOG, AUTO-READ, AUTOSTIR, Auto-Test, BOD AutoEZ, Cable-Free, CERTI-CAL, CISA, DataCOLLECT, DataPLUS, digital LogR, DirectCal, DuraProbe, Environmental Product Authority, Extra Easy/Extra Value, FAST QC, GAP, GLPcal, GLPcheck, GLPdoc, ISEasy, KAP, LabConnect, LogR, Low Maintenance Triode, Minimum Stir Requirement, MSR, NISS, One-Touch, One-Touch Calibration, One-Touch Measurement, Optimum Results, Orion Star, Pentrode, pHuture MMS, pHuture Pentrode, pHuture Quatrode, pHuture Triode, Quatrode, QuiKcheK, rf link, ROSS Resolution, SAOB, SMART AVERAGING, Smart CheK, SMART STABILITY, Stacked, Star Navigator 21, Stat Face, The Enhanced Lab, ThermaSense, Triode, TRIUMpH, Unbreakable pH, Universal Access are trademarks of Thermo Fisher.
Guaranteed Success and The Technical Edge are service marks of Thermo Fisher.
PerpHecT meters are protected by U. S. PerpHecT ROSS electrodes are protected by U. S. ORION Series A meters and 900A printer are protected by U. S. [. . . ] (Refer Section 5. 4 for Buffer Selection Settings) The factory default is USA standard. It is recommended that you perform 2-point calibration at room temperature (25 oC), starting with the first buffer at pH 6. 86 followed by any other buffer value. (pH 4. 01, 1. 68, 9. 18 or 12. 45)
- 19 -
1-Point Calibration:
1
CAL CAL
2
CAL
pH
3
mV
4
MEAS
ENT
HOLD
READY HOLD
CAL
HOLD
ENT
READY
PH
pH
pH
°C ATC
1 2
From pH measurement mode press CAL key to enter calibration mode as described in section 4. 2. Immerse the temperature probe in the buffer solution if ATC mode is enabled. Immerse the potential matching pin in the buffer if symmetrical mode is enabled. Lower display shows pH 6. 86 (NIST standard buffer). Upper display shows the current uncalibrated pH reading. LCD shows `READY' annunciator when the reading is stable. The pH Monitor re-calculates electrode properties based on the calibration. The new slope (in mV) is shown in the upper display and pH reading at 0mV (pH7. 00 ± offset) is shown in the lower display. Press ENT key to exit from the calibration. The pH Monitor reverts to pH measurement mode.
3
4
2-Point Calibration:
1
CAL
2
CAL CAL
pH
3
CAL
pH
4
mV
5
MEAS
ENT
READY
ENT
READY
ENT
ENT
READY
PH
pH
pH
CAL
pH
°C ATC
3
CAL
READY pH
ENT
pH
CAL
3
CAL
READY pH
ENT
pH
CAL
3
CAL
READY pH
ENT
pH
CAL
- 20 -
1 2
From pH measurement mode press CAL key to enter calibration mode as described in section 4. 2. Press ENT key to begin first calibration point. Immerse the temperature probe in the buffer solution if ATC mode is enabled. Immerse the potential matching pin in the buffer if symmetrical mode is enabled. Lower display shows pH 6. 86 (NIST standard buffer). Upper display shows the current uncalibrated pH reading. LCD shows `READY' annunciator when the reading is stable. The pH Monitor moves to the second calibration point. The lower display shows next standard buffer value (pH 4. 01). Use and keys to select your second buffer from one of the preset values: pH 4. 01 or 1. 68 or 9. 18 or 12. 45. Remove the electrode from the first buffer, wash and then immerse it into the second buffer. [. . . ] pH De-ionized water rinse sachets, 20 ml x 20 pcs pH sachet assortment pack 5 each of pH 4. 01, pH 7. 00, pH 10. 01 and de-ionized water sachets per box. Protein cleaning solution for pH electrode Storage solution for pH electrode Order Code EC-BU-4BT EC-BU-7BT EC-BU-10BT EC-BU-4BS EC-BU-7BS EC-BU-10BS EC-RIN-WT EC-AST-PK EC-DPC-BT EC-RE-005 Order Code TSPHCP0550 EC-ARTSO-05B EC-ARGTSO-05B EC-100GTSO-05B EC-DA93506-05B EC-HTAUTSO-05B EC-HTPTTSO-05B
NOTE: · · · pH buffer solutions (480 ml bottle) have ±0. 01 pH accuracy at 25 °C pH buffer sachets are individually sealed, single use pouch containing 20 ml of fresh, contamination free calibration solution pH buffer sachets have ±0. 01 pH accuracy at 25°C
- 33 -
7. 2
Eutech Instruments Order Code
pH Monitor Replacement and Accessories
Item Description Alpha pH 550 Monitor pH/Temp electrode with PMP and 10-ft cable Platinum ORP electrode with 10-ft cable BNC to spade lug adapter Order Code 56717-30 35807-20 35801-21 05994-90
Calibration Solutions
Item Description pH 4. 01 calibration buffer, 500 ml pH 7. 01 calibration buffer, 500 ml pH 10. 01 calibration buffer, 500 ml pH 4. 01 calibration buffer solution pouches, 20/box pH 7. 00 calibration buffer solution pouches, 20/box pH 10. 00 calibration buffer solution pouches, 20/box pH De-ionized water rinse sachets, 20 ml x 20 pcs Assortment pack 5 each of pH 4. 01, pH 7. 00 and pH 10. 00 solution pouches. Electrode cleaning solution Electrode storage solution Order Code 00654-00 00654-04 00654-08 35653-01 35653-02 35653-03 35653-00 35653-04 00653-06 00653-04
NOTE: · To order Eutech accessories, contact the nearest Oakton distributor
- 34 -
8
TROUBLESHOOTING
Cause a) b) Loose connections Incorrect output voltage of the power adaptor Solution a) b) a) Ensure cables make good contact Use an power adaptor with specified output voltage
Problem Power on, but no display
Unstable pH reading
Oscillating temperature readings Slow response Blinking ATC
Dirty electrode Electrical noise interference a) Electrical noise interference a) Dirty / Oily electrode a) No temperature probe connection during ATC mode Error in calibration
a) b)
Clean electrode and recalibrate b) Switch to Symmetric mode a) Ensure shield wire is properly connected to pin 7 a) Clean electrode a) Ensure temperature sensing cable makes good contact Ensure calibration standard solution is not contaminated. Ensure electrode makes good contact with Transmitter Ensure electrode makes good contact with Transmitter
Blinking electrode annunciator Or (pH)
a)
a)
a)
pH electrode is not connected Temperature probe is not connected when ATC enabled
a)
Or (Temperature)
a)
a)
- 35 -
9
9. 1
GENERAL INFORMATION
Warranty
This monitor is supplied with a one-year warranty against significant deviations in material and workmanship from date of purchase and a six-month warranty for probe. [. . . ]