If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Lastmanuals provides you a fast and easy access to the user manual EUTECH INSTRUMENTS CYANIDE EPOXY. We hope that this EUTECH INSTRUMENTS CYANIDE EPOXY user guide will be useful to you.
Lastmanuals help download the user guide EUTECH INSTRUMENTS CYANIDE EPOXY.
Manual abstract: user guide EUTECH INSTRUMENTS CYANIDE EPOXY
Detailed instructions for use are in the User's Guide.
[. . . ] Instruction Manual
Cyanide Electrode
EUTECH INSTRUMENTS PTE LTD.
INSTRUCTION MANUAL
CYANIDE ION ELECTRODE
1
Cyanide Electrode
Instruction Manual
TABLE OF CONTENTS
GENERAL INSTRUCTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Required Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Required Solutions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
GENERAL PREPARATION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Electrode Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Electrode Slope Check (with pH/mV meter) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Electrode Slope Check (with an ion meter). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
MEASUREMENT. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Measuring Hints . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Sample Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Units of Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
MEASUREMENT PROCEDURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . [. . . ] A new calibration curve should be prepared daily.
4.
5.
6.
7.
Direct Measurement of Cyanide (using an ion meter)
1.
By serial dilution of the 1. 0X10-2M or 1, 000 ppm cyanide standard, prepare two standards whose concentration is near the expected sample concentration. Measure 100 ml of each standard into individual 150 ml beakers and add 1 ml of 10M NaOH (ISA) to each. Place the more dilute solution on the magnetic stirrer and begin stirring at a constant rate. Assure that the meter is in the concentration mode. Lower the electrode tips into the solution. Adjust the meter to the concentration of the cyanide standard and fix the value in the memory according to the meter manufacturer's instructions after stabilization of the reading. Rinse the electrodes with distilled water and blot dry. Place the more concentrated solution on the magnetic stirrer and begin stirring at a constant rate. Lower the electrode tips into the solution. Adjust the meter to the concentration of the cyanide standard and fix the value in the memory according to the meter manufacturer's instructions after stabilization of the reading.
7
2.
3. 4.
5.
6.
Cyanide Electrode
Instruction Manual
7.
For low level measurements (below 8X10-6M or 0. 2 ppm), place the rinsed, dried electrodes into a solution containing 100 ml of distilled water and 1 ml ISA. After stabilization, fix the blank value in the meter according to the meter manufacturer's instruction. After rinsing the electrodes and blotting dry, place the electrode tips into 100 ml of the sample and 1 ml of ISA. After stabilization, read the concentration directly from the meter display. The calibration should be checked every 2 hours. Assuming no change in ambient temperature, place the electrode tips in the first cyanide standard. After the reading has stabilized, compare it to the original reading in Step 4 above. A reading differing by more than 0. 5 mV or a change in the ambient temperature will necessitate the repetition of Step 2-10 above. The meter should be re-calibrated daily.
8.
9.
Low-Level Cyanide Measurements (using a pH/mV meter)
Use the following low level cyanide measurement procedure in the non-linear portion of the calibration curve (below 8X10-6M or 0. 2 ppm). A more accurate electrode indicator technique, such as titration, using a silver/sulfide ion electrode, may be preferred below these levels. By serial dilution, prepare 100 ml of 1. 0X10-3M or 10 ppm cyanide standard from the 1. 0X10-2M or 1, 000 ppm cyanide standard. Using a 150 ml plastic beaker, add 100 ml of distilled water and 1 ml ISA. [. . . ] Millivolt potentials occur from the inter-diffusion of ions in the two solutions. Electrode charge will be carried unequally across the solution boundary resulting in a potential difference between the two solutions, since
13
Cyanide Electrode
Instruction Manual
ions diffuse at different rates. When making measurements, it is important to remember that this potential be the same when the reference is in the standardizing solution as well as in the sample solution or the change in liquid junction potential will appear as an error in the measured electrode potential. The composition of the liquid junction filling solution in the reference electrode is most important. [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE EUTECH INSTRUMENTS CYANIDE EPOXY
Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service.