If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Lastmanuals provides you a fast and easy access to the user manual EUTECH INSTRUMENTS CYBERSCAN CON 510 CONDUCTIVITY METER. We hope that this EUTECH INSTRUMENTS CYBERSCAN CON 510 CONDUCTIVITY METER user guide will be useful to you.
Lastmanuals help download the user guide EUTECH INSTRUMENTS CYBERSCAN CON 510 CONDUCTIVITY METER.
Manual abstract: user guide EUTECH INSTRUMENTS CYBERSCAN CON 510 CONDUCTIVITY METER
Detailed instructions for use are in the User's Guide.
[. . . ] Instruction Manual
CON 510
Bench Conductivity/TDS Meter
68X090820 Rev 2 01/04
Technology Made Easy . . .
Preface
This manual serves to explain the use of the CON 510 bench meter. It functions in two ways, firstly as a step by step guide to help you to operate the meter. Secondly, it serves as a handy reference guide. It is written to cover as many anticipated applications of the meter as possible. [. . . ] Taking measurements with the Auto-Hold feature selected on When a reading has stabilized for more than 5 seconds, the Auto-Hold feature automatically "freezes" the displayed reading and the [HOLD] indicator appears. You may deactivate the Auto-Hold feature as described in SETUP sub-menu P3. 1.
11
Instruction Manual
CON 510
5. 4
Using Manual Ranging Function
ENTER RANGE
Although the CON 510 bench meter has an automatic ranging capability, you may also manually select any specific measurement range you wish to work on by pressing RANGE each time for every range. Conductivity Range 0 20. 00 µS/cm 0 200. 0 µS/cm 0 2000 µS/cm 0 20. 00 mS/cm 0 200. 0 mS/cm TDS Range (if TDS factor is 0. 5) 0 10. 00 ppm 0 100. 0 ppm 0 1000 ppm 0 10. 00 ppt 0 100 ppt
MEAS
READY
uS
.
ENTER RANGE
o
C
ATC
MEAS
uS
o
2. 3.
To manually select the desired measuring range, press RANGE key while you are in the measurement mode. The first range will appear on the display and the [MEAS] indicator blinks. Press RANGE key again (if needed) until the desired range is selected. To re-select the Auto-ranging function, repeatedly press RANGE key until the [MEAS] indicator appears without blinking. The display will then scroll through the various ranges and the Auto-ranging function resumes. Or simply power off and on again will restart meter into Auto-ranging function as default.
C
ATC
ENTER RANGE
MEAS
READY
uS
.
o
C
ATC
NOTES: If the value of the solution you are measuring is higher than the range selected, an error message code "Or" will appear on the upper display indicating an over-range condition. Press RANGE key until the correct range is selected. However the meter reverts back to the Auto-ranging function once it is powered off.
5. 5
HOLD Function
HOLD
This feature allows you to freeze the displayed reading momentarily when activated in the measurement mode and the [HOLD] indicator appears on the LCD. To release the held value, press HOLD again and the [HOLD] indicator disappears. You may continue to take measurements.
MEAS
READY HOLD
uS
.
o
C
ATC
12
Instruction Manual
CON 510
6
MEMORY AND DATA INPUT FUNCTIONS
6. 1
Memory Input
MEAS
READY
The meter's non-volatile memory can store and recall up to 50 data sets which include Conductivity and temperature or TDS and temperature. Data sets are retained even if the power is shut off unless these stored data sets are being overwritten by new ones. During any measurement mode press MI/ key to input any data into the memory. MEM, "Sto" and memory location number will be displayed to indicate that the readings are being stored into meter's memory at that specific location number. The meter then returns to the measurement mode. To continue storing data sets, press MI/ to input subsequent data into memory. NOTE: If the memory is full, the first data set value stored will be overwritten to create space for the new value. Data sets are stored sequentially in the next unoccupied memory location and cannot be selectively input into specific memory location to be stored.
uS
.
MEM
o
C
ATC
6. 2
Memory Recall
ENTER RANGE
You can access memory recall from any measurement mode only. This function recalls and displays previous readings being stored in the meter's memory in "Last-In-First-Out" sequence. For example the most recent reading stored in location 20 will be displayed first when memory recall is activated. [. . . ] The result should be in TDS value.
27
Instruction Manual
CON 510
15 ADDENDUM 3: CALCULATING TEMPERATURE COEFFICIENTS
To determine the temperature coefficient of your sample solution use this formula:
Where: tc = Temperature coefficient CT1 = Conductivity at Temp 1 T1 = Temp 1 25 = 25 °C CT2 = Conductivity at Temp 2 T2 = Temp 2
NOTE: A controlled temperature water bath is ideal for this procedure. Immerse the probe into a sample solution and adjust the temperature coefficient to 0% (that is, no compensation) by performing the following: A. From measurement mode, press the SETUP key to enter into [SETUP] mode. Use MI/ or MR/ key until the lower display shows "P5. 0". [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE EUTECH INSTRUMENTS CYBERSCAN CON 510 CONDUCTIVITY METER
Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service.