If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Lastmanuals provides you a fast and easy access to the user manual EUTECH INSTRUMENTS CYBERSCAN DO 1500 DO METER. We hope that this EUTECH INSTRUMENTS CYBERSCAN DO 1500 DO METER user guide will be useful to you.
Lastmanuals help download the user guide EUTECH INSTRUMENTS CYBERSCAN DO 1500 DO METER.
Manual abstract: user guide EUTECH INSTRUMENTS CYBERSCAN DO 1500 DO METER
Detailed instructions for use are in the User's Guide.
[. . . ] Instruction Manual
CyberScan DO 1500
Dissolved Oxygen Meter
Technology Made Easy . . .
68X292336 Rev. 1 01/04
PREFACE
Thank you for selecting the Eutech Instruments CyberScan DO 1500 bench meter. The instruction manual serves to explain the use of the CyberScan DO 1500 bench meter as a step-by-step operational guide to help you familiarise with the meter's features and functions. It is structured sequentially with illustration of diagrams that explains the various functions and setup menus available. [. . . ] 4.
Press
or
key to toggle between Curr or Store.
Press enter key to confirm selection and to go to next setup page OR press std to exit from this page without confirming the change mode.
P rint Print
P rint Print
30
Instruction Manual
CyberScan DO 1500
4. 1. 12.
P12. 0: Clear Stored Data
This setup option allows you to clear all stored data sets (from previous measurements) in the meter's memory for new data to be stored. Note old data sets will be overwritten by any new data sets in the event when the stored locations have exceeded.
P rint clear
press to set value pres s to s value et pres enter to accept s press enter to accept
To Clear Stored Data 1. Access the Setup pages menu from the measurement mode by pressing the setup key. Press the or key in setup mode to access the Clear Stored Data setup page. Press enter to access the setup page.
31
Instruction Manual
CyberScan DO 1500
3. 4.
Press
or
key to toggle between YES or NO.
Press enter key to confirm selection and to go to next setup page OR press std to exit from this page without confirming the change mode.
When you press enter key to confirm the clearing of memory, meter displays blinking "clear" wait until meter display stop blinking to proceed with next button press.
32
Instruction Manual
CyberScan DO 1500
4. 1. 13.
P13. 0: Clear Buffer Values
This setup option allows you to clear the standardised buffer values.
mg/L
% Sat
clear BUFFER
press
to set value
press enter to accept
To Clear Buffer Values 1. Access the Setup pages menu from the measurement mode by pressing the setup key. Press the or key in setup mode to access the Clear Buffer Values setup page. Press enter to access the setup page.
33
Instruction Manual
CyberScan DO 1500
3. Press
or
key to toggle between YES or NO.
4. Press enter key to confirm selection and to go to next setup page OR press std to exit from this page without confirming the change mode.
If you are doing user reset from %sat setup page, it will clear both % sat and mg/L calibration. If you are doing user reset from mg/L setup page, meter will clear mg/L calibration only.
34
Instruction Manual
CyberScan DO 1500
5. %Sat One Point calibration (Auto)
User can do the 100%Sat calibration. The pressure that has been set in the setup will be applied only during the calibration. Accepted window for 100% sat= 50% - 200%Sat 1. Press std to enter to `Standardize' screen. Upper display shows the present measured value, while lower display shows the pressure compensated 100% sat. Wait for the upper display value to stabilise. Meter blinks the cal values for few seconds before exiting to measure screen.
Standardize
% Sat
press enter to accept
35
Instruction Manual
CyberScan DO 1500
5. 2.
%Sat Two Point calibration (Auto)
User must do the 0. 0%Sat cal first. This point will be considered as offset. After successful 0%Sat calibration, meter will prompt for the 100% Sat calibration. Otherwise meter displays the error message. [. . . ] CLEANING
This meter requires no regular maintenance, but it is recommended to occasionally wipe down the front with a damp cloth from time to time.
11. TROUBLESHOOTING
The CyberScan DO 1500 displays pertinent error messages to guide you should an error occur with a measurement or meter operation.
Message Error Icon "---" Or Ur Err1
Description Error message for %Sat Cal error. DO reading is out of range Over range condition for temp Under range condition for temp Communication error
45
Instruction Manual
CyberScan DO 1500
12. ACCESSORIES
Consult your Authorised Distributors for these items.
12. 1.
Replacement Meters and Accessories
DESCRIPTION CyberScan DO 1500 Bench DO/BOD Meter with self stirring BOD probe (EC-620-SSP), Data Acquisition Software and 110VAC power adapter, 2-flat pin US type (center negative) CyberScan DO 1500 Bench DO/BOD Meter with self stirring BOD probe (EC-620-SSP), Data Acquisition Software and 220VAC power adapter, 2-round pin EURO type (center negative) 6 membrane caps, polishing disk and electrolyte solution. [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE EUTECH INSTRUMENTS CYBERSCAN DO 1500 DO METER
Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service.