If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Lastmanuals provides you a fast and easy access to the user manual EUTECH INSTRUMENTS DO 6. We hope that this EUTECH INSTRUMENTS DO 6 user guide will be useful to you.
Lastmanuals help download the user guide EUTECH INSTRUMENTS DO 6.
Manual abstract: user guide EUTECH INSTRUMENTS DO 6REV 2
Detailed instructions for use are in the User's Guide.
[. . . ] Instruction Manual
DO 6
Economy Hand-held Dissolved Oxygen Meter
68X243622 Rev 2 Apr 08
Technology Made Easy . . .
Part of Thermo Fisher Scientific
Preface
This manual serves to explain the use of the DO 6 Dissolved Oxygen Handheld. This manual functions in two ways: first as a step by step guide to help you operate the meter; second, as a handy reference guide. This manual is written to cover as many anticipated applications of the DO6 palm-top meter as possible. If there are doubts in the use of the Dissolved Oxygen Handheld, please do not hesitate to contact the nearest Eutech Instruments / Oakton Instruments authorized Distributor. [. . . ] HOLD can be used in any measurement mode. 1) To hold a measurement, press the HOLD/ENTER key while in measurement mode. The %, mg/L, ppm or T annunciator in the display will start blinking to indicate that the measurement is being held. To release the held value, press the HOLD/ENTER again. The annunciator will stop blinking to indicate that the held value has been released. Continue to take measurements.
2)
3)
A
%
Annunciator Blinks when meter is in "HOLD" mode
Figure 17: Unit in HOLD mode
Note: This meter has a selectable auto off feature which when activated will shut the meter off automatically after 20 minutes of non use. If the meter is shut off either automatically or manually, the HOLD value will be lost
30
Instruction Manual
DO 6
7
ADVANCED SETUP FUNCTIONS
7. 1
Advanced Setup Overview
The advanced setup mode lets you customize your meter's preferences and defaults. There are two Set Up menus: - one which is derived from the % Saturation or Temperature mode and the other is derived from the mg/L (ppm) Concentration mode. 7. 1. 1 1) 2) 3) 4) 7. 1. 2 5) 6) To enter the % Saturation or Temperature Set Up menu: Make sure that the meter is switched-off while in the % Saturation or Temperature measurement mode. With the MODE key pressed, power the unit on by pressing the ON key. Release the ON key before releasing the MODE key. "SEt. P" indicator will appear momentarily and "COF. 1" will appear next. You are now in the % Saturation or Temperature Set Up menu. To enter the mg/L (ppm) Concentration Set Up menu: Switch off the meter in the mg/L (ppm) Concentration measurement mode. Repeat steps 2 to 3 as above and you will be in the mg/L (ppm) Concentration Set Up menu.
The following table combines the explanation for the features available in both Set Up menus. Figures 19 and 20 give an overall view of the Set Up sequence for the % Saturation or Temperature Set Up menu and the mg/L (ppm) Concentration Set Up menu.
31
Instruction Manual SEt. P Enter Set Up Menu
DO 6
COF. 1 Enter Configuration Menu A. ATC Selection for automatic temperature compensation. b. DO Selection of mg/L or ppm of the concentration mode measurement. (Only available in Concentration Set Up menu) CAL. 2 Viewing the latest Calibration data `View only' parameter for the latest calibration data of the DO % Saturation calibration and the DO mg/L (ppm) Concentration calibration. Calibration data will be viewed according to its respective Set Up menu. ELE. 3 Viewing the Electrode properties FACT Viewing the Slope Factor OFS Viewing the % Saturation offset adjustment. (Only available in % Saturation Set Up menu) HI. mV Viewing the mV value for 100% Saturation LO. mV Viewing the mV value for 0% Saturation AtO. 4 Selection of automatic power off (20 minutes from the last key press)
rSt. 5 Enter Reset Menu A. CAL Selection of Calibration Reset (Clear only Calibration and its Data back to factory default condition) b. USR Selection of User Reset (Clear all user setting back to factory default condition)
32
Instruction Manual
DO 6 OFS. 6 - % Saturation Offset Adjustment (Only available in % Saturation Set Up menu)
DPr. 7 Enter Barometric Pressure Selection Menu A. HG Selection of Barometric Pressure Setting Adjustment in mmHg A. PA - Selection of Barometric Pressure Setting Adjustment in Kilo Pascal b. SAL Enter the Salinity Setting Adjustment Menu (Only available in mg/L (ppm) Concentration Set Up menu)
33
Instruction Manual
Switch off from % Saturation or Temperature measurement mode. Press and hold MODE key and then switch on.
DO 6
MODE
ON OFF
SEt. P will display for 1 second
Figure 18: Overview of DO 6 % Saturation or Temperature Setup Menu
34
Instruction Manual
MODE
ON OFF
DO 6
Switch off from Concentration measurement mode. [. . . ] 14. 1. 5 Applications
Oxygen is essential for fish, invertebrate, plant, and aerobic bacteria respiration. DO levels below 3 ppm are stressful to most aquatic organisms. Levels below 2 or 1 ppm will not support fish. Fish growth and activity usually require 5 to 6 ppm of DO, an important consideration for Aqua-culture industry. [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE EUTECH INSTRUMENTS DO 6
Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service.