If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Lastmanuals provides you a fast and easy access to the user manual EUTECH INSTRUMENTS SILVERSULFIDE EPOXY. We hope that this EUTECH INSTRUMENTS SILVERSULFIDE EPOXY user guide will be useful to you.
Lastmanuals help download the user guide EUTECH INSTRUMENTS SILVERSULFIDE EPOXY.
Manual abstract: user guide EUTECH INSTRUMENTS SILVERSULFIDE EPOXY
Detailed instructions for use are in the User's Guide.
[. . . ] Instruction Manual
Silver/Silver Sulfide Electrode
EUTECH INSTRUMENTS PTE LTD.
INSTRUCTION MANUAL
SILVER/SILVER SULFIDE ION ELECTRODE
1
Silver/Silver Sulfide Electrode
Instruction Manual
TABLE OF CONTENTS
GENERAL INSTRUCTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Required Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Required Solutions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . [. . . ] Place 100 ml of the sample and 2 ml of ISA in a 150 ml beaker, place it on the magnetic stirrer, and begin stirring. Immerse the electrode tips in the solution and wait for the reading to stabilize. Read the concentration directly from the meter display. The calibration should be checked every two hours. Assuming no change in ambient temperature, place the electrode tips in the first silver standard. After the reading has stabilized, compare it to the original reading in Step 4 above. A reading differing by more than 0. 5 mV or a change in the ambient temperature will necessitate the repetition of Steps 2-8 (9) above. The meter should be re-calibrated daily.
5. 8.
9.
10.
11.
12.
Direct Measurement of Sulfide (using a pH/mV meter)
1.
By serial dilution of the weekly standard, prepare three standard sulfide solutions. Measure out 50 ml of each standard into individual 150 ml beakers and add 25 ml of SAOB and 25 ml of distilled water to each. Place the most dilute solution on the magnetic stirrer and begin stirring at a constant rate. After assuring that the meter is in the mV mode, lower the electrode tips into the solution. When the reading has stabilized, record the mV reading. Place the midrange sulfide standard on the magnetic stirrer and begin stirring. After rinsing the electrodes with distilled water, blot dry and immerse the electrode tips in the solution. When the reading has stabilized, record the mV reading.
2.
3.
10
Instruction Manual
Silver/Silver Sulfide Electrode
4.
Place the most concentrated sulfide standard on the magnetic stirrer and begin stirring. After rinsing the electrodes with distilled water, blot dry and immerse the electrode tips in the solution. When the reading has stabilized, record the mV reading. Using the semi-logarithmic graph paper, plot the mV reading (linear axis) against concentration (log axis). A typical calibration curve can be found in Figure 1.
5.
A calibration curve is constructed on semi-logarithmic paper when using a pH/mV meter in the millivolt mode. The measured electrode potential in mV (linear axis) is plotted against the standard concentration (log axis). In the linear region of the curve, only three standards are necessary to determine a calibration curve. [. . . ] Millivolt potentials occur from the inter-diffusion of ions in the two solutions. Electrode charge will be carried unequally across the solution boundary resulting in a potential difference between the two solutions, since ions diffuse at different rates. When making measurements, it is important to remember that this potential be the same when the reference is in the standardizing solution as well as in the sample solution or the change in liquid junction potential will appear as an error in the measured electrode potential.
21
Silver/Silver Sulfide Electrode
Instruction Manual
The composition of the liquid junction filling solution in the reference electrode is most important. The speed with which the positive and negative ions in the filling solutions diffuse into the sample should be as nearly equal as possible, that is, the filling solution should be equitransferent. [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE EUTECH INSTRUMENTS SILVERSULFIDE EPOXY
Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service.
Click on "Download the user manual" at the end of this Contract if you accept its terms, the downloading of the manual EUTECH INSTRUMENTS SILVERSULFIDE EPOXY will begin.