[. . . ] RPM4/HPMS A70M/A20M-AFTM Reference Pressure Monitor with
High Pressure Mounting System Operation and Maintenance Manual
NSN 6685-01-529-6990 RN (2 of 3)
© 2007 DH Instruments, a Fluke Company
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL
High pressure liquids and gases are potentially hazardous. Energy stored in these liquids and gases can be released unexpectedly and with extreme force. High pressure systems should be assembled and operated only by personnel who have been instructed in proper safety practices.
This instrument is not to be operated in any other manner than that specified by the manufacturer.
© 2007 DH Instruments, a Fluke Company
All rights reserved.
Information in this document is subject to change without notice. No part of this document may be reproduced or transmitted in any form or by any means, electronic or mechanical, for any purpose, without the express written permission of DH Instruments, a Fluke Company 4765 East Beautiful Lane, Phoenix AZ 85044-5318, USA. [. . . ] These functions are considered part of RPM4 maintenance and are covered in the maintenance section of this manual (see Sections 5. 2, 5. 3).
3. 5. 9
<9RESET>
PURPOSE To reset various RPM4 settings to default or factory values. PRINCIPLE RPM4 stores its user definable settings in non-volatile memory. The reset menu allows the user to selectively or completely reset these settings to factory defaults. This clears any settings that the user has made, and should be used only to restore the RPM4 to a known state. RPM4 goes through its power up sequence after any type of reset is executed. RPM4 reset functions will change current settings to factory defaults. These may include settings vital to RPM4 operation and affecting the calibration of the quartz reference pressure transducers (Q-RPTs). Reset functions should only be used by qualified personnel with knowledge of reset consequences. Reset functions should never be used "experimentally". OPERATION To access the RESET menu, press [SPECIAL], <9reset>. The display is: RESET menu choices include: <1set> <2units> to reset general system operating parameters (see Section 3. 5. 9. 1). to reset unit of measure functions (see Section 3. 5. 9. 2). 1sets 2units 3ATest 4cal 5all
<3ATest> to reset AutoTest parameters, clear File AutoTest files and clear the AutoTest data log (see Section 3. 5. 9. 3). <4cal> <5all> to reset internal calibration coefficients and modes (see Section 3. 5. 9. 4). to reset all settings except ID and security password to factory default values (see Section 3. 5. 9. 5).
3. 5. 9. 1
<1SETS>
PURPOSE
Sets most general operating parameters back to default values. Does not affect calibration coefficients, remote interfaces or AutoRange ranges. The Reset Sets resets are itemized in Table 9.
Page 69
© 2007 DH Instruments, a Fluke Company
RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL
Table 9. Reset Sets
RESET [UNIT] [MODE] [RANGE] [HEAD] Stability Limit Upper Limit Resolution AutoZ AutoZ Leak Check Screen Saver Key Sounds Lo Vnt ReadRate RESULT Pressure unit of measure to first of the six available (psi) Measurement mode to the native mode of the Q-RPT (absolute) Hi Q-RPT default range (10000 psi) 0 cm height and Nitrogen medium 0. 005 % of Hi Q-RPT default range Hi Q-RPT default range default value 0. 001 % FS of Hi Q-RPT default range AutoZ ON for both Q-RPTs and gauge and asbsolute measurement modes. Zoffset value not affected Patm, 0 set to 101. 325 kPa a 15 second run time. Clear logged results 10 minutes to activation Medium tone valid key press sound Automatic Automatic SEE SECTION 3. 3. 2 3. 3. 3 3. 3. 1 3. 3. 7 3. 4. 3 3. 4. 4 3. 4. 2 3. 5. 1 3. 5. 1, 3. 2. 2 3. 3. 5 3. 5. 5. 1 3. 5. 5. 2 3. 5. 7. 4 3. 5. 7. 2
3. 5. 9. 2
<2 UNITS>
PURPOSE
Sets the pressure units available under the UNIT function to default selections depending (see Section 3. 3. 2). Sets the user defined unit to 1. 000/Pa (see Section 3. 5. 6). Sets the reference temperature for inWa unit to 20°C.
3. 5. 9. 3
· · ·
<3ATEST>
PURPOSE
Resets Auto Tests to default characteristics (see Section 3. 3. 10). [. . . ] This creates a pseudo Q-RPT designated HL. If the pressure measured by the Q-RPT exceeds Pmax!, an overpressure condition occurs. Pressure controller calibrator manufactured by DHI. RPM4 can be used as an external reference device for a PPC3. [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE FLUKE A70M Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service. Click on "Download the user manual" at the end of this Contract if you accept its terms, the downloading of the manual FLUKE A70M will begin.