If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Lastmanuals provides you a fast and easy access to the user manual FLUKE MOLSTIC HI FLOW. We hope that this FLUKE MOLSTIC HI FLOW user guide will be useful to you.
Lastmanuals help download the user guide FLUKE MOLSTIC HI FLOW.
Manual abstract: user guide FLUKE MOLSTIC HI FLOWINSTALLATION
Detailed instructions for use are in the User's Guide.
[. . . ] Ideally, inlet supply pressure is regulated to approximately 760 kPa (110 psig). In no case should the supply exceed 1 MPa (150 psig). Note: Be sure the molstic isolation valve (1) is closed (handle down) before connecting the quick connector stem to the quick connector. [. . . ] If desired, install the DUT platform (2) to assure that the DUT is at the correct height for alignment with the molstic downstream fitting (see DUT PLATFORM INSTALLATION AND OPERATION below). If you have a metering valve option for manual flow control, install it here. The metering valve option requires use of the 1/2 in. adaptor.
Q U IC K C O N N E C T S TE M W ITH ¼ IN . S W AG EL OK FITTIN G QU IC K C O N N E C TO R
½ IN . VCR
molsticTM, hi flow, molblocTM Mounting System Instruction Sheet
molbloc INSTALLATION
!Loosen the interconnecting nipple clamp (3) so that the nipple can move freely in its rubber grommet. Move the nipple away from the molbloc cradles to allow maximum room for molbloc insertion. # Place the molbloc in the molstic molbloc cradles (be sure flow orientation is correct). $ Connect and tighten the upstream (valve side) molbloc VCR connection using a soft O-ring and following the procedure provided in Document 560009 (Recommendations for molbloc Installation) supplied with the molbloc. % Pull the interconnecting nipple towards the molbloc downstream connection. Connect and tighten the downstream molbloc VCR connection using a soft O-ring and following the procedure provided in Document 560009 (Recommendations for molbloc Installation) supplied with the molbloc. & Tighten the interconnecting nipple clamp (3) until the nipple is secure (5 N/m max). ' Connect the molbloc to the molbox and leak check the system. ( To remove a molbloc from the molstic, reverse the procedure. Break the downstream molbloc VCR connector and pull the interconnecting nipple away from the bloc. Break the upstream molbloc VCR connector. Remove the molbloc.
Q U IC K C O N N E C T S TE M W ITH ¼ IN . [. . . ] Nupro is a registered trademark of NUPRO Co. SWG and VCR are registered trademarks of The Swagelok Companies. Document 560040a 010626
)
DH Instruments, Inc.
4765 East Beautiful Lane Phoenix AZ 85044-5318 USA
Calibration Solutions for Pressure and FlowTM
Tel 602. 431. 9100 Fax 602. 431. 9559 dhi@dhinstruments. com www. dhinstruments. com
molbloc, molbox and molstic are trademarks of DH Instruments, Inc. [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE FLUKE MOLSTIC HI FLOW
Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service.