Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... DON'T FORGET : ALWAYS READ THE USER GUIDE BEFORE BUYING !!!
If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Lastmanuals provides you a fast and easy access to the user manual HANNA INSTRUMENTS HI 8614L. We hope that this HANNA INSTRUMENTS HI 8614L user guide will be useful to you.
Lastmanuals help download the user guide HANNA INSTRUMENTS HI 8614L.
You may also download the following manuals related to this product:
HANNA INSTRUMENTS HI 8614L (575 ko)
Manual abstract: user guide HANNA INSTRUMENTS HI 8614L
Detailed instructions for use are in the User's Guide.
[. . . ] Instruction Manual
HI 8614 - HI 8614L HI 8615 - HI 8615L
pH and ORP Transmitters
pH
pH INDIC ITTER SM & TRAN
L HI 8614OR AT
These Instruments are in Compliance with the CE Directives
http://www. hannainst. com
1
Dear Customer, Thank you for choosing a Hanna Instruments Product. Please read this instruction manual carefully before using the instrument. This manual will provide you with all the necessary information for the correct use of the instrument, as well as a precise idea of its versatility in a wide range of applications. These instruments are in compliance with the directives.
TABLE OF CONTENTS Preliminary Examination . . . . . . . . . . . . . . . . . . . . . . . . . . . . [. . . ] If routine check is required, the following procedures shall be performed. · Follow the procedure at page 7 to perform calibration on the transmitter module (using a ammeter). Simulate a 4. 00 mA loop current for the transmitter (i. e. pH 0. 00 as Electrode input) and check for display reading. Simulate a 20. 00 mA loop current for the transmitter (i. e. pH 14. 00 as Electrode input) and check for display reading.
·
·
Note: when the instrument is used in conjunction with the Hanna indicator HI 8510T or controllers HI 8710T, HI 8711T, the calibration can also be performed on the indicator/controller. In this case slight calibration adjustment can be made on the indicator/ controller even if the whole system calibration is advised, always starting from the transmitter.
12
pH CALIBRATION WITH FIXED TEMPERATURE COMPENSATION (HI 8614 & HI 8614L only) · Take the temperature of the buffer solutions using a Checktemp or a thermometer with a resolution of at least 1°. Connect the appropriate resistor to the "TEMP. PROBE" terminals (see page 6) depending on the temperature of the calibration solution. Follow the procedure as outlined in the calibration with automatic temperature compensation and use the thermometer reading to adjust the trimmer until the ammeter reads the value according to the following table.
4. 01 pH Buffer 7. 01 pH Buffer 10. 01 pH Buffer
°C
·
·
Temperature
°C 0 5 10 15 20 25 30 35 40 45 50 55 60
°F 32 41 50 59 68 77 86 95 104 113 122 131 140
pH 4. 01 4. 00 4. 00 4. 00 4. 00 4. 01 4. 02 4. 03 4. 04 4. 05 4. 06 4. 07 4. 09
mA 8. 576 8. 565 8. 565 8. 565 8. 565 8. 576 8. 587 8. 599 8. 610 8. 622 8. 633 8. 645 8. 667
pH 7. 13 7. 10 7. 07 7. 04 7. 03 7. 01 7. 00 6. 99 6. 98 6. 98 6. 98 6. 98 6. 98
mA 12. 137 12. 103 12. 069 12. 034 12. 023 12. 000 11. 989 11. 977 11. 966 11. 966 11. 966 11. 966 11. 966
pH 10. 32 10. 24 10. 18 10. 12 10. 06 10. 01 9. 96 9. 92 9. 88 9. 85 9. 82 9. 79 9. 77
mA 15. 789 15. 697 15. 629 15. 560 15. 491 15. 434 15. 377 15. 331 15. 285 15. 251 15. 217 15. 182 15. 159
13
ORP CALIBRATION (HI8615 & HI8615L) Initial preparation: Disconnect the +ve supply cable from the "+VE LOOP" terminal and connect a 20 mA f. s. ammeter between the +ve cable and the "+VE LOOP" terminal. With HI 8615L the instrument display can be used during calibration without the need to connect the ammeter. In this case the values are directly expressed in mV units. Pour a small quantity of HI 7020 ORP solution into a beaker. Procedure: · Connect the shorting BNC connector to the ORP transmitter. · Adjust the OFFSET ADJUSTMENT trimmer on the module for a display of 12 mA on the ammeter or 0 mV on the instrument display (HI 8615L only). This sets the zero point for the transmitter.
20 70 HI
SLOPE ADJUSTMENT OFFSET ADJUSTMENT
mA
·
Connect the ORP electrode to the transmitter and immerse the tip of the electrode into a beaker of HI 7020 ORP calibration solution, and check that the ammeter reading lies between 13. 6 and 14. 2 mA or the instrument reading is between 200 and 275 mV at 25°C (HI 8615L).
mA
HI 7020
14
If the reading lies outside this range, adjust the slope adjustment trimmer on the transmitter for a reading just within this range. The unit is now calibrated.
·
SLOPE ADJUSTMENT OFFSET ADJUSTMENT
A complete calibration of the transmitter module is advised periodically. This calibration procedure requires the HI 8427 or the HI 931001 pH and ORP simulator to simulate the ORP electrode. HI 8427 or HI 931001 produce a known signal into the system so that the faults of the system can be isolated. DO NOT BE ALARMED IF ANY SALT DEPOSITS ARE PRESENT. This is normal with electrodes and they will disappear when rinsed with water. [. . . ] It takes longer time for the reading to stabilize if the resistance is higher. In addition, the response time will suffer to a greater degree at temperatures below 10°C.
2x109 -10 1x109
-10 -10 2x108 1x108 -10
2x107 -10 1x107
-10
-20 -20 0 +10+20 +30+40 +50+60+70 +80+90 ° C
Since the resistance of the pH electrode is in the range of 200 Mohm, the current across the membrane is in the pico Ampere range. Large currents can disturb the calibration of the electrode for many hours.
For these reasons high humidity environments, short circuits and static discharges are detrimental for a stable pH reading.
The pH electrode's life also depends on the temperature. If constantly used at high temperatures, the electrode life is drastically reduced.
23
Typical Electrode Life Ambient Temperature 1- 3 years 90 °C Less than 4 months 120°C Less than 1 month
High concentrations of sodium ions interfere with readings in alkaline solutions; the pH at which the interference starts to be significant depends upon the composition of the glass. [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE HANNA INSTRUMENTS HI 8614L
Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service.
Click on "Download the user manual" at the end of this Contract if you accept its terms, the downloading of the manual HANNA INSTRUMENTS HI 8614L will begin.