If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Lastmanuals provides you a fast and easy access to the user manual HANNA INSTRUMENTS HI 8751. We hope that this HANNA INSTRUMENTS HI 8751 user guide will be useful to you.
Lastmanuals help download the user guide HANNA INSTRUMENTS HI 8751.
You may also download the following manuals related to this product:
HANNA INSTRUMENTS HI 8751 (509 ko)
HANNA INSTRUMENTS HI 8751 BROCHURE (509 ko)
Manual abstract: user guide HANNA INSTRUMENTS HI 8751
Detailed instructions for use are in the User's Guide.
[. . . ] These instruments are in compliance with the directives EN 50081-1 and EN 50082-1.
TABLE OF CONTENTS
PRELIMINARY EXAMINATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 GENERAL DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 8751 . . . . . . . . . 3 FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 8752 . . . . . . . . . [. . . ] HI 8751 measures temperature in Centigrade whereas HI 8752 reads temperature in Fahrenheit and HI 8753 has both ranges available for measurement. Each meter comes supplied with the precalibrated and interchangeable HI 765BL temperature probe, a 9V battery and an instruction manual.
2
FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 8751
1) Probe connector 2) LCD display 3) OFF mode 4) °C scale 5) Rotary switch
SPECIFICATIONS Range Resolution Accuracy (@20°C/68°F) EMC Typical Deviation Display Battery Probe (included) Environment Dimensions Weight HI 8751 40. 0 to 150. 0°C 0. 1°C ±0. 5% F. S. , for one year, excluding probe error ±1°C, with HI 765BL probe 3½ digit LCD 9V battery providing 300 hours of continuous operation HI 765BL precalibrated and interchangeable stainless steel probe with 1 m (3. 3') cable 0 to 50°C (32 to 122°F); max 95% RH non-condensing 180 x 83 x 40 mm (7. 1x3. 3x1. 6") 280 g (10 oz. ) with probe
3
FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 8752
1) Probe connector 2) LCD display 3) OFF mode 4) °F scale 5) Rotary switch
SPECIFICATIONS Range Resolution Accuracy (@20°C/68°F) EMC Typical Deviation Display Battery Probe (included) Environment Dimensions Weight HI 8752 58 to 338°F 1°F ±0. 5% F. S. , for one year, excluding probe error ±2°F, with HI 765BL probe 3½ digit LCD 9V battery providing 300 hours of continuous operation HI 765BL precalibrated and interchangeable stainless steel probe with 1 m (3. 3') cable 0 to 50°C (32 to 122°F); max 95% RH non-condensing 180 x 83 x 40 mm (7. 1x3. 3x1. 6") 280 g (10 oz. ) with probe
4
FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 8753
1) Probe connector 2) LCD display 3) OFF mode 4) °C scale 5) °F scale 6) Rotary switch
SPECIFICATIONS Range Resolution Accuracy (@20°C/68°F) EMC Typical Deviation Display Battery Probe (included) Environment Dimensions Weight HI 8753 40. 0 to 150. 0°C / 58 to 338°F 0. 1°C / 1°F ±0. 5% F. S. , for one year, excluding probe error ±3°C / ±6°F, with HI 765BL probe 3½ digit LCD with symbols 9V battery providing 300 hours of continuous operation HI 765BL precalibrated and interchangeable stainless steel probe with 1 m (3. 3') cable 0 to 50°C (32 to 122°F); max 95% RH non-condensing 180 x 83 x 40 mm (7. 1x3. 3x1. 6") 280 g (10 oz. ) with probe
5
OPERATIONAL GUIDE
INITIAL PREPARATION
Remove the battery cover from the back of your thermometer. Unwrap the battery and connect it to the clips. Insert the battery and replace the cover. Connect the temperature probe to the probe socket on the top of the meter. To switch on, turn the rotary switch to °C or °F (depending on model).
OFF
°C °F
If a temperature probe is plugged in, the unit will display the temperature. After use, switch your thermometer off by turning the rotary switch to OFF. OFF °C
°F
CALIBRATION
All Hanna Instruments thermometers have been accurately precalibrated at the factory. It is generally recommended to have all thermometers recalibrated at least once a year. For an accurate annual recalibration, contact your nearest Hanna Customer Service Center. Hanna Test Plugs provide a quick and easy way to test the accuracy of the meter by simply connecting these Test Plugs to the probe connector on the instrument. If the reading differs by more than ±0. 4°C (±0. 8°F) from the Test Plug value, the unit is due for recalibration.
6
Choose the right Test Plug to best suit your needs: HI 765-18C Calibration key, 18. 0°C ±0. 4°C HI 765000C Calibration key, 0. 0°C ±0. 4°C HI 765070C Calibration key, +70. 0°C ±0. 4°C HI 765-004F Calibration key, 0. 4°F ±0. 8°F HI 765032F Calibration key, +32. 0°F ±0. 8°F HI 765158F Calibration key, +158. 0°F ±0. 8°F
BATTERY REPLACEMENT
The instrument is powered by a 9V battery located at the back of the instrument. When the battery is rundown, the Battery Error Preventive System (BEPS) circuitry is activated. BEPS shuts down the LCD automatically to ensure that no erroneous readings are taken due to a weak battery. In such cases, the battery needs to be replaced for the correct functioning of the instrument. To access the battery, remove the battery cover by applying pressure in the direction indicated. Unplug the rundown battery and replace it with a new one. Re-insert the battery and replace the cover. Battery replacement must only take place in a non-hazardous area using a 9V alkaline battery.
BATTERY
7
TEMPERATURE PROBES & ACCESSORIES
Hanna Instruments offers a wide range of probes to meet your needs to measure the temperature in air, surface, penetration and liquid applications. These probes use highly sensitive thermistor sensors which provide greater accuracy, faster response and a temperature range that is superior to conventional thermistor probes. [. . . ] Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc. , 584 Park East Drive, Woonsocket, Rhode Island, 02895 , USA.
Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
10
CE DECLARATION OF CONFORMITY
DECLARATION OF CONFORMITY
We Hanna Instruments Srl V. le delle industrie 12 35010 Ronchi di Villafranca (PD) ITALY herewith certify that the thermometers
HI 8751
IEC 801-2 IEC 801-3 EN 55022
HI 8752
HI 8753
have been tested and found to be in compliance with the following regulations: Electrostatic Discharge RF Radiated Radiated, Class B
Date of Issue: 23-11-1995 D. Volpato - Engineering Manager On behalf of Hanna Instruments S. r. l.
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environment in which they are used. Operation of these instruments in residential areas could cause unacceptable interference to radio and TV equipment. Any variation introduced by the user to the supplied equipment may degrade the instruments' EMC performance. To avoid electrical shock, do not use these instruments when voltage at the measurement surface exceeds 24VAC or 60VDC. [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE HANNA INSTRUMENTS HI 8751
Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service.
Click on "Download the user manual" at the end of this Contract if you accept its terms, the downloading of the manual HANNA INSTRUMENTS HI 8751 will begin.