If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Lastmanuals provides you a fast and easy access to the user manual HANNA INSTRUMENTS HI 9224. We hope that this HANNA INSTRUMENTS HI 9224 user guide will be useful to you.
Lastmanuals help download the user guide HANNA INSTRUMENTS HI 9224.
Manual abstract: user guide HANNA INSTRUMENTS HI 9224
Detailed instructions for use are in the User's Guide.
[. . . ] Instruction Manual
HI9224 - HI92240
Portable Microprocessor Printing and Logging pH Meters
instru m
n e ts
pH
MICR pH M OPROC ETER ESSO R PR INTI
HI 92
24
NG
ON/O FF
PAP E
R
CAL
CFM
BUF F
RAN
GE
PRIN T
LOG
http://www. hannainst. com
This Instrument is in Compliance with the CE Directives
Dear Customer, Thank you for choosing a Hanna Instruments Product. Please read this instruction manual carefully before using the instrument. This manual will provide you with all the necessary information for the correct use of the instrument, as well as a precise idea of its versatility in a wide range of applications. These instruments are in compliance with directives EN 500811 and EN 50082-1.
TABLE OF CONTENTS
PRELIMINARY EXAMINATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . [. . . ] The log number and page number will appear for a few seconds on the display to indicate the current operational mode. The instrument will print a complete set of data and the "LOG" symbol will appear on the bottom left hand-side of the LCD.
ALT
CAL
CON
LOG
Log number
Page number
RANGE
TIME
LOG
The printout provides the following information: a - Date (DD-MM-YY) b - A running log number c - A running sample number (in that particular log) d - Printing interval indicator in minutes 01-07-95 ==01== *0001M b e - time (HH-MM) 0001 11. 01 c 7. 06 pH 17. 5 °C f - pH value g - temperature value
19
a d e f g
To log without printing press the ALT and the PAPER keys at the same time and the "LOG" symbol on display will start to blink. Press ALT and PAPER again to resume logging with printing.
PRINT
PAPER
LOG
ALT
CAL
After approximately 5 minutes the display will switch off but the logging function remains active. While logging, during the sleep mode, the display will periodically flash on. This is a normal indication that the meter is in the logging mode and is currently logging readings. To reactivate the display press the RANGE key. Note Once in the logging mode, the interval cannot be changed. Exit the logging mode first (pressing the ALT ALT and the LOG keys) before setting the RANGE new interval. Note If the PRINT key is pressed while in logging mode, a printout is produced without affecting the running sample number.
RANGE
CAL
CON
TIME
LOG
PRINT
SAMPLE NUMBER
During logging it is possible to know the running sample number. Press the LOG key twice and the display will show the number of values that have been taken since the start of the current log.
LOG
LOGGING MODE WITH PRINTING
This function is suggested when an immediate report of the measurement is required in addition to the recording of the data into memory. Press the ALT and the LOG key simultaneously to enter the logging mode. The log CAL CON ALT number and page number will appear for a few seconds on the display to indicate the RANGE TIME LOG correct operational mode. The printer will
20
print a complete set of data and the "LOG" symbol will appear on the bottom left hand side of the display. If no key is pressed, the display goes blank after about 5 minutes. During printing, the display shows the time, interval and the "LOG" symbol.
LOG
To reactivate the display press the RANGE key. The printout provides the following information: a - Date (DD-MM-YY) b - A running log number c - A running sample number (in that particular log) d - Printing interval indicator in minutes 01-07-95 e - Time (HH-MM) ==01== *0001M b 0001 11. 01 c 7. 06 pH f - pH value 17. 5 °C g - Temperature value
RANGE
a d e f g
It is always possible to switch from the logging with printing function to the logging function. Press the ALT and the PAPER keys at the same time and the "LOG" symbol will start to blink to indicate that the data is now stored only into memory.
PRINT
PAPER
ALT
CAL
Notes: · It is recommended to use the adapter during logging in printing mode, especially when many printouts are going to be taken. · Before proceeding with logging with printing, make sure there is enough paper for your measurements. When the paper is finished the meter will not advise the operator and the printouts could be lost. If this happens, data will continue to be stored in memory, and it is always possible to print the
21
data at different time (see above). · It is possible to insert a new paper roll during logging session (see page 27) · Once in the logging mode, the interval cannot be changed. Exit the logging mode first (pressing the ALT and the LOG keys) before setting the new interval. [. . . ] Use: milk, yogurt.
12 mm 0. 5"
FC 210 120 mm 4. 7"
FC 220B Glass-body, triple-ceramic, single junction, refillable, combination pH electrode. Use: food processing.
9. 5mm DIA 0. 37" 12 mm 0. 5"
FC 220 120 mm 4. 7"
Kynar® is registered Trademark of "Pennwalt Corp. " Ultem® is registered Trademark of "General Electrics Co. "
38
FC 911B Plastic-body (Kynar®), double junction, refillable with built-in amplifier, combination pH electrode. Use: very high humidity.
12 mm 0. 5"
FC 911
110 mm 4. 3"
HI 1413B Glass-body, single junction, flat tip, Viscolene, non-refillable, combination pH electrode. Use: surface measurement.
12 mm 0. 5"
HI 1413
110 mm 4. 3"
EXTENSION CABLES FOR SCREW-TYPE ELECTRODES ONLY (SCREW TO BNC CONNECTOR):
HI7855/1 HI7855/3 HI7855/5 HI7855/10 HI7855/15 Extension cable 1m (3. 3') long Extension cable 3m (9. 9') long Extension cable 5m (16. 5') long Extension cable 10m (33') long Extension cable 15m (49. 5') long
HI 7855 SERIES CABLE CONNECTORS CONNECTOR AND 3. 0 mm (0. 12") CABLE WITH BNC
CONNECT TO SCREW TYPE ELECTRODES
CONNECT TO THE BNC SOCKET OF THE METER
Kynar® is registered Trademark of "Pennwalt Corp. "
39
POWER UNITS:
HI710005 HI710006 Voltage adapter from 115 VAC to 12 VDC Voltage adapter from 230 VAC to 12 VDC
OTHER ACCESSORIES
Rugged carrying case Plain Paper Spare Rolls (10 pcs) Spare Ink Cartridge (1 pc) 1. 5V AA batteries (10 pcs) Electrode holder Temperature probe with 1 m (3. 3') screened cable pH and ORP electrode simulator with 1 m (3. 3') coaxial cable ending in female BNC connectors (HI 7858/1) HI 9200 Infrared Transmitter HI 92000/16 Windows® 3. 11 compatible software for data transfer to PC HI 92000/32 Windows® 95 compatible software for data transfer to PC HI 931001 pH and ORP electrode simulator with LCD display 1 m (3. 3') coaxial cable ending in female BNC connectors (HI 7858/1) MANPHPRNR2Instruction manual HI 710031 HI 710034 HI 710035 HI 721308 HI 76405 HI 7669/2W HI 8427
Windows® is registered Trademark of "Microsoft Co. "
40
ELECTRODE APPLICATION REFERENCE GUIDE
Application 1. [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE HANNA INSTRUMENTS HI 9224
Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service.