If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Lastmanuals provides you a fast and easy access to the user manual HANNA INSTRUMENTS HI 93716. We hope that this HANNA INSTRUMENTS HI 93716 user guide will be useful to you.
Lastmanuals help download the user guide HANNA INSTRUMENTS HI 93716.
You may also download the following manuals related to this product:
HANNA INSTRUMENTS HI 93716 (491 ko)
Manual abstract: user guide HANNA INSTRUMENTS HI 93716
Detailed instructions for use are in the User's Guide.
[. . . ] Insert a sample before adding reagent and press ZERO. A blinking "0. 00" indicates that the sample absorbs less light than the zero reference. Check the procedure and make sure you use the same cuvet for reference (zero) and measurement. A flashing value higher than the maximum concentration readable (see specifications) indicates that the sample absorbs too much light, meaning that the concentration is too high. [. . . ] If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Customer Service department and then send it with shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection. To validate your warranty, fill out and return the enclosed warranty card within 14 days from the date of purchase. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc. , Woonsocket, Rhode Island, 02895 , USA. Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
The HI 93716 meter measures the bromine content in water and wastewater in the 0. 00 to 8. 00 mg/L (ppm) range. The meter uses an exclusive positive-locking system to ensure that the cuvet is in the same place every time it is placed into the measurement cell. The reagents are in powder form and are supplied in packets. The amount of reagent is precisely dosed to ensure maximum repeatability. Display codes aid the user in routine operations. The meters have an auto-shut off feature that will turn the instrument off after 10 minutes of non-use.
Light Emitting Diode @ 555 nm Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18 t h edition, DPD method. The reaction between bromine and the reagent causes a pink tint in the sample Light Detector Silicon Photocell Environment 0 to 50°C (32 to 122°F); max 95% RH non-condensing Battery Type/Life 1 x 9 volt/40 hours Auto-Shut off After 10' of non-use Dimensions 180 x 83 x 46 mm (7. 1 x 3. 3 x 1. 8") Weight 290 g (10 oz. ).
REQUIRED REAGENTS
Code HI 93716-0 Description DPD Reagent Quantity 1 packet
V
REAGENT SETS
HI 93716-01 Reagents for 100 tests HI 93716-03 Reagents for 300 tests
OPERATIONAL GUIDE
MEASUREMENT PROCEDURE
· Turn the meter on by pressing ON/OFF.
ON OFF
· Press READ TIMED and the display will show the countdown prior to the measurement, or, alternatively wait for 2 minutes and 30 seconds and press READ DIRECT. In both cases "SIP" will appear during measurement.
12 9
TIPS FOR AN ACCURATE MEASUREMENT
The instruction listed below should be carefully followed during testing to ensure best accuracy. · Do not touch the cuvet walls with hands. · In order to maintain the same conditions during the zeroing and the measuring phases, it is necessary to close the cuvet to prevent any contamination. · Do not let the test sample stand too long after reagent is added or accuracy will be lost. · Whenever the cuvet is placed into the measurement cell, it must be completely free of fingerprints, oil or dirt. Wipe it thoroughly with HI 731318 or a lint-free cloth prior to insertion. [. . . ] In case of water with hardness greater than 500 mg/L CaCO3, shake the sample for approximately 1 minute after adding the reagent.
CE DECLARATION OF CONFORMITY
· Wait for a few seconds and the display will show "-0. 0-". Now the meter is zeroed and ready for measurement.
· Remove the cuvet and add the content of one packet of HI 93716 reagent. Replace the cap and shake gently.
BATTERY REPLACEMENT
Battery replacement must only take place in a non-hazardous area using a 9V alkaline battery. Simply slide off the battery cover on the back of the meter. [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE HANNA INSTRUMENTS HI 93716
Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service.