If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Lastmanuals provides you a fast and easy access to the user manual HANNA INSTRUMENTS HI 955201. We hope that this HANNA INSTRUMENTS HI 955201 user guide will be useful to you.
Lastmanuals help download the user guide HANNA INSTRUMENTS HI 955201.
You may also download the following manuals related to this product:
HANNA INSTRUMENTS HI 955201 (653 ko)
Manual abstract: user guide HANNA INSTRUMENTS HI 955201
Detailed instructions for use are in the User's Guide.
[. . . ] Untitled-10
1
2
Instruction Manual
HI 955201 - HI 955202 HI 955301 - HI 955302
Portable Microprocessor Printing and Logging PT-100 Thermometers
Dear Customer, Thank you for choosing a Hanna Instruments Product. Please read this instruction manual carefully before using the instrument. This manual will provide you with all the necessary information for the correct use of the instrument, as well as a precise idea of its versatility in a wide range of applications. These instruments are in compliance with directives EN 50081-1 and EN 50082-1. [. . . ] The printout provides the following information: a 10-01-96 d ==07== 0005M b 0014 *14. 25 c e a - Date (DD-MM-YY) #1 25. 0 °C. b - A running log number c - Running sample number (in that particular lot) d - Printing interval indicator in minutes e - Time (HH-MM) f - Temperature value/s
19
LOGGING MODE WITH PRINTING This function is suggested when an immediate report of the measurement is required in addition to the recording of the data into memory. Press the ALT and LOG keys simultaneously to enter the logging mode. The log number and page number will appear for a few seconds on the display to indicate the correct operational mode. The printer will print a complete set of data and the "LOG" symbol will appear on the bottom left corner of the display.
ALT UP DOWN TEMP TIME LOG
LOG
20
Untitled-10
21
22
If no key is pressed, the display goes blank after about 5 minutes. During printing, the display shows the time, interval and the "LOG" symbol.
·
TIME
LOG
INTV
· ·
To reactivate the display press the TEMP key.
TEMP
The printout provides the following information: a - Date (DD-MM-YY) b - A running log number c - A running sample number (in that particular log) d - Printing interval indicator in minutes e - Time (HH-MM) f - Temperature value/s g - Channel # (HI 955302 only) It is always possible to switch from the logging with printing function to the logging only function. Press the ALT and the PAPER keys at the same time and the "LOG" symbol will start to blink to indicate that the data are now stored only into memory.
b c 10-01-96 ==07== 0005M 0014 *14. 25 #1 25. 0 °C. a d e f
°C
·
Before proceeding with logging and printing, make sure there is enough paper for your measurements. When the paper is finished the meter will not advise the operator and the printouts could be lost. If this happens, data will continue to be stored into memory, and it is always possible to print them at different time (see above). It is possible to insert a new paper roll during logging session (see page 31) Once in the logging mode, the interval cannot be changed. Exit the logging mode first (pressing the ALT and the LOG keys) before setting the new interval. If the PRINT key is pressed while in logging mode, a printout is produced without affecting the PRINT running sample number.
ALT UP DOWN TEMP TIME LOG
TO STOP LOGGING Press ALT and LOG keys simultaneously, this will generate a log exit status printout.
ALT
UP
DOWN
TEMP
TIME
LOG
Log number Total sample number
==07== 0016
10-01-96 0005M *16. 25
LOG EXIT Date/Time
PRINT
PAPER
LOG
CH
ALT
UP
Notes: · It is recommended to use the adapter during logging with printing mode, especially when many printouts are going to be taken.
21 22
Untitled-10
23
24
TO SCAN STORED DATA ON DISPLAY
Press the LOG key. The display will show the log number and the page number of the next logging.
Press UP and the temperature will be displayed.
°C
CH
LOG
While pressing the ALT key, press DOWN until the log number to scan appears on the secondary display. The primary display will show the number of samples of that particular log.
Continue pressing UP to display all the memorized data of the same log one by one in the above sequence i. e. Press the DOWN key to revert back to sampling time and move back through samples. To exit from the recall mode press LOG.
ALT
DOWN
LOG
Note: this mode will not alter data into memory Press ALT and TEMP simultaneously. This now shows the date on which the log started.
ALT
DATE
TEMP
Press UP and the time will be displayed.
TIME
UP
TO PRINT STORED DATA Once a log number is selected by using the ALT and DOWN, as detailed in the chapter "TO SCAN STORED DATA ON DISPLAY" you can print all or part of that log section by using the ALT and PRINT keys. The printer will start to print the logged section beginning with the selected sample number without altering the content of the memory. Note: It is always possible to print only the sample shown on the display by pressing PRINT.
ALT
DOWN
PRINT
ALT
PRINT
23
24
Untitled-10
25
26
Note: Before proceeding with printing, make sure there is enough paper for the data to be printed. When the paper is finished the meter will not advise the operator and printouts could be lost. If this happens, stop the printer by pressing ALT PAPER and PAPER simulta- PRINT neously. [. . . ] Replace the battery cover and secure screws.
-
1. 5V 1. 5V 1. 5V 1. 5V
+
+
+
-
+
-
SCREW POTS RESET
Insert the paper edge in the printer slot and feed the printer by pressing the PAPER key.
PAPER
33
34
Untitled-10
35
36
BATTERY REPLACEMENT If "LO BAT" appears on the display, it is an indication that the batteries are running down. If it appears during printing, it means that 200 printouts can be made before the batteries are exhausted. When there is only sufficient power for 100 printouts, the "LO BAT" sign is displayed continuously on the LCD.
°C
CH
Note: The instrument uses the following configuration.
+
-
LOBAT
Battery replacement must only take place in a non hazardous area using the battery types specified in this instruction manual (see page 35).
It is recommendable to purchase Hanna Instruments HI 710005 and HI 710006 voltage adapters that use the proper polarity configuration. HI 955201, HI 955202, HI 955301 and HI 955302 can also be used with other adapters. [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE HANNA INSTRUMENTS HI 955201
Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service.