If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Lastmanuals provides you a fast and easy access to the user manual HANNA INSTRUMENTS HI 983317-0. We hope that this HANNA INSTRUMENTS HI 983317-0 user guide will be useful to you.
Lastmanuals help download the user guide HANNA INSTRUMENTS HI 983317-0.
Manual abstract: user guide HANNA INSTRUMENTS HI 983317-0
Detailed instructions for use are in the User's Guide.
[. . . ] Any defective item must be returned in its original packing.
Trimmers CAL For calibration SET To adjust setpoint LEDs SET MEAS ALARM
Front view of the panel-mounted unit
CAL
42mm 1. 65" 49mm 1. 93"
73mm 2. 87"
79mm 3. 11"
SET
MEAS
SET
ON when LCD displays the set value ON when LCD displays the measured value ON when alarm contact is activated
REAR PANEL
1. Power Supply: HI 983317-0 HI 98331700 12 VDC HI 983317-1 115/230 VAC
WARRANTY
These meters are warranted for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. The probe is warranted for a period of six months. This warranty is limited to repair or replacement free of charge. [. . . ] Measurements are highly accurate and the meters can be calibrated at one point. Power supply, wiring and selection are made via the plugin terminal blocks on the rear panel. LED indicators on the front panel identify whether the controller is in set or measurement mode, and dosing mode.
Side view of the panel-mounted unit Supplied adjustable location brackets allow the controller to slide into the cutout and will hold the unit securely in place. 95 mm (3. 74") is the minimum amount of space required to install the controller.
0. 25/4mm 0. 01/0. 160"
ADJUSTABLE LOCATION BRACKET
79mm 3. 11"
76mm 3" 95mm MIN 3. 74"
L1: Positive 230 VAC L2: Negative 115 VAC L3: -------Neutral L4: --------Protection Earth Note: the power input is internally protected by a 400 mA fuse 2. This contact acts as a switch for the power. The contact has to be protected outside by the user.
Panel-mounted unit assembling view
HI 740146
Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
OPERATIONAL GUIDE
POWER CONNECTION HI 983317-0 and HI 98331700 Connect a 2-wire power cable to the terminal strip while paying attention to the correct positive and negative polarities (12 VDC). HI 983317-1 Connect a 3-wire power cable to the terminal strip while paying attention to the correct earth, neutral and line contacts (115/230 VAC). Notes: · All external cables connected to the rear panel should end with wire lugs. · It is recommended to cover the unused terminals with insulating tape. · At start up the meter needs a few seconds to stabilize. Wait until a stable reading is displayed. Any initial variation on readings may be due to temperature compensation. The "ALARM" LED will light up when the alarm contact is closed, to indicate an EC value lower than selected setpoint. CALIBRATION Make sure the meter is in the measurement mode (the "MEAS" LED lights on). Immerse the probe in HI 7039 (5. 00 mS/cm) calibration
solution.
ACCESSORIES
HI 7632-00 EC probe HI 70039P 5. 00 mS/cm (5000 µS/cm) calibration solution, 20mL sachet (25 pcs) HI 7039M 5. 00 mS/cm (5000 µS/cm) calibration solution, 230mL bottle HI 7061M Electrode cleaning solution, 230 mL bottle HI 710005 12 VDC power adapter, US plug HI 710006 12 VDC power adapter, European plug HI 710012 12 VDC power adapter, Australian plug HI 710013 12 VDC power adapter, Southern Africa plug HI 710014 12 VDC power adapter, UK plug HI 731326 Calibration Screwdriver (20 pcs) HI 740146 Mounting brackets
CE DECLARATION OF CONFORMITY
Shake briefly and wait for reading to stabilize. Using a small screwdriver adjust the calibration trimmer until the meter displays "5. 00" on the LCD. The display will show the default or previously adjusted value for the setpoint. Using a small screwdriver, adjust the "SET" trimmer until the required limit value is displayed. PROBE MAINTENANCE
SPECIFICATIONS
HI 983317
Range 0. 00 to 10. 00 mS/cm Resolution 0. 01 mS/cm Accuracy (@ 20°C) ±2% f. s. Setpoint Adjustable through multiturn trimmer Alarm Condition LED ON and alarm contact closed when measured EC value is lower than setpoint Alarm Output 2-contact relay, no fuse protected. [. . . ] · When not in use, clean the probe before storing it.
Recommendations for Users Before using these products, make sure that they are entirely suitable for the environment in which they are used. Operation of these instruments in residential areas could cause unacceptable interferences to radio and TV equipment. The metal band at the end of the probe is sensitive to electrostatic discharges. Avoid touching this metal band at all times. [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE HANNA INSTRUMENTS HI 983317-0
Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service.