If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Lastmanuals provides you a fast and easy access to the user manual HANNA INSTRUMENTS HI 7635. We hope that this HANNA INSTRUMENTS HI 7635 user guide will be useful to you.
Lastmanuals help download the user guide HANNA INSTRUMENTS HI 7635.
Manual abstract: user guide HANNA INSTRUMENTS HI 7635
Detailed instructions for use are in the User's Guide.
[. . . ] Instruction Manual
GENERAL DESCRIPTION
Dear Customer, Thank you for choosing a Hanna product. This manual will provide you with the necessary information for the correct use of the probe. [. . . ] Do not use in systems where the temperature exceeds 80°C (176°F).
INSTALLATION
For the installation of the in-line conductivity probes, it is necessary to use rubber seals between the probe and the pipe or elbow joints. A pipe sealant is also recommended to ensure a leak free joint. When screwing the joints, take care not to overtighten as excessive pressures can cause the probe damage.
HI 7635 - HI 7636
4-PIN IN-LINE CONDUCTIVITY PROBES
TYPICAL IN-LINE ASSEMBLY FOR HI 7635 (or HI 7636).
www. hannainst. com
Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
CONNECTIONS
The probes are supplied with a 4 m cable. The wires from the cable must be connected to the HI 8936 process Conductivity Transmitter as shown below. The connections are color coded for an easy installation. BLACK RED GREY/BROWN BLUE WHITE GREEN NTC SENSOR (HI 7635 only) NTC SENSOR (HI 7635 only) probe pin 1 probe pin 2 probe pin 3 probe pin 4
PROBE MAINTENANCE
The probes can be compensated for normal contamination by a process of re-calibration. However, it is recommended to remove regularly the process conductivity probes from the system for maintenance. Deposits on the probe can be removed by immersing it in 0. 1N Hydrochloric acid for about 30 minutes. Heavier deposits may demand longer immersion time. Clean the electrode thoroughly with water prior to the reinstallation. On reinstalling, check the seals carefully to ensure a leak connection. Note: Always recalibrate the meter when attaching a new probe.
WARRANTY
All Hanna Instruments probes are warranted for a period of six months. [. . . ] When shipping any instrument, make sure it is properly packaged for complete protection.
TYPICAL CONNECTION OF THE HI 7635 (or HI 7636) PROBE AND THE HI 8936 TRANSMITTER TO THE HI 8931 CONTROLLER.
MECHANICAL DIMENSIONS
142mm 5. 60"
52mm 2. 04"
73mm 2. 87"
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner.
1" NPT
IST7635R1 06/05 [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE HANNA INSTRUMENTS HI 7635
Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service.