If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Lastmanuals provides you a fast and easy access to the user manual HANNA INSTRUMENTS HI 8633. We hope that this HANNA INSTRUMENTS HI 8633 user guide will be useful to you.
Lastmanuals help download the user guide HANNA INSTRUMENTS HI 8633.
Manual abstract: user guide HANNA INSTRUMENTS HI 8633
Detailed instructions for use are in the User's Guide.
[. . . ] Instruction Manual
HI 8633 HI 8733 - HI 8734
Reliable and Waterproof Multi-Range Conductivity Meters
www. hannainst. com
Dear Customer, Thank you for choosing a Hanna Product. Please read this instruction manual carefully before using the instrument. It will provide you with the necessary information for a correct use of the instrument, as well as a more precise idea of its versatility. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst. com. [. . . ] @25°C, 1382 mg/L.
°C
HI
9
· All subsequent measurements will be compensated to 25°C (77°F). If you prefer to standardize the temperature compensation to 20°C (68°F) rather than 25°C (77°F), leave the TEMPERATURE knob at 18°C (if the temperature of the solution is 18°C), adjust the trimmer to read "11. 67 mS" (see the conductivity vs. temperature chart) or "1251 mg/L" (see the TDS vs. All subsequent measurements will be compensated to 20°C. · The calibration is now complete and the instrument is ready for use. The instrument should be recalibrated at least once a month, or when the probe is changed. Note: For more accurate results, it is advisable to use a calibration solution close to the measurement range. See the "Accessories" section for a wide selection of conductivity solutions.
· Remove the battery compartment cover and adjust the calibration trimmer located inside with the calibration screwdriver until the display shows "12. 88 mS" i. e. the conductivity reading @ 25°C.
PROCEDURE FOR HI 8733
· Pour sufficient quantity of a conductivity calibration solution (e. g. HI 7030/ HI 8030) into a beaker to cover the holes on the probe. If possible, use plastic beakers to minimize any EMC interference. · Immerse the conductivity probe in the solution, making sure that holes are completely submerged. · Wait for a couple of minutes for thermal equilibrium to be reached. · Tap the probe on the bottom, then shake it while rotating to make sure no air bubbles remain trapped in the sleeve. · Switch the instrument on by pressing ON/OFF. · Set the temperature coefficient knob to 2% to compensate for the temperature effect of average solutions (to determine exact value for a particular solution, see page 14). · Select 19. 99 mS/cm range by pressing the appropriate range key. 10
ON OFF
HI 703 0
· All subsequent measurements will be compensated to 25°C (77°F). If you prefer to standardize the temperature compensation to 20°C (68°F) rather than 25°C (77°F), adjust the trimmer to read "11. 67 mS" (see the conductivity vs. All subsequent measurements will be compensated to 20°C. · The calibration is now complete and the instrument is ready for use. The instrument should be recalibrated at least once a month, or when the probe is changed. [. . . ] When shipping any instrument, make sure it is properly packaged for complete protection.
CE DECLARATION OF CONFORMITY
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner.
Recommendations for Users Before using these products, make sure that they are entirely suitable for the environment in which they are used. Operation of these instruments in residential area could cause unacceptable interference to radio and TV equipment, requiring the operator to take all necessary steps to correct interferences. The metal band at the end of the probe is sensitive to electrostatic discharges. [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE HANNA INSTRUMENTS HI 8633
Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service.