If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Lastmanuals provides you a fast and easy access to the user manual HANNA INSTRUMENTS HI 93701. We hope that this HANNA INSTRUMENTS HI 93701 user guide will be useful to you.
Lastmanuals help download the user guide HANNA INSTRUMENTS HI 93701.
You may also download the following manuals related to this product:
HANNA INSTRUMENTS HI 93701 (404 ko)
Manual abstract: user guide HANNA INSTRUMENTS HI 93701
Detailed instructions for use are in the User's Guide.
[. . . ] A flashing value higher than the maximum readable concentration (see specifications) indicates that the sample absorbs too much light, meaning that the concentration is too high. The cuvet is not inserted correctly and an excess ambient light is reaching the detector. If the cover is properly installed, then contact your dealer or the nearest Hanna Customer Service Center. The zero sample is too dark for proper zeroing. [. . . ] When shipping any instrument, make sure it is properly packaged for complete protection. To validate your warranty, fill out and return the enclosed warranty card within 14 days from the date of purchase. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc. , Woonsocket, Rhode Island, 02895 , USA. Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
The HI 93701 portable microprocessor meter measures the free chlorine (Cl2) content in water and wastewater in the 0. 00 to 2. 50 mg/L (ppm) range. The meter uses an exclusive positive-locking system to ensure that the cuvet is in the same place every time it is placed into the measurement cell. The reagents are in liquid or powder form and are supplied in bottles or in packets. The amount of reagent is precisely dosed to ensure maximum repeatability. Display codes aid the user in routine operations. The meters have an auto-shut off feature that will turn the instrument off after 10 minutes of non-use.
±0. 01 mg/L Light Emitting Diode @ 555 nm Adaptation of the EPA recommended DPD method 330. 5. The reaction between free chlorine and the DPD reagent causes a pink tint in the sample Light Detector Silicon Photocell Environment 0 to 50°C (32 to 122°F); max 95% RH non-condensing Battery Type/Life 1 x 9 volt/40 hours Auto-Shut off After 10' of non-use Dimensions 80 x 83 x 46 mm (7. 1 x 3. 3 x 1. 8") Weight 290 g (10 oz. )
REQUIRED REAGENTS
POWDER: Code HI 93701-0 LIQUID: Code HI 93701A-F HI 93701B-F Description DPD Description DPD1 Indicator DPD1 Buffer Quantity 1 packet Quantity 3 drops 3 drops
Operational Guide
MEASUREMENT PROCEDURE
· Turn the meter on by pressing ON/OFF. · When the LCD displays "- - -", it is ready. · Fill the cuvet with 10 mL of 10 mL unreacted sample, up to the mark, and replace the cap. · Place the cuvet into the holder and ensure that the notch on the cap is positioned securely into the groove. · Press ZERO and "SIP" will appear on the display.
ZERO
· The instrument directly displays concentration in mg/L of free chlorine on the Liquid Crystal Display.
ON OFF
Tips for an Accurate Measurement
The instruction listed below should be carefully followed during testing to ensure best accuracy. · Do not touch the cuvet walls with hands. · In order to maintain the same conditions during the zeroing and the measuring phases, it is necessary to close the cuvet to prevent any contamination. · Do not let the test sample stand too long after reagent is added or accuracy will be lost. · Whenever the cuvet is placed into the measurement cell, it must be completely free of fingerprints, oil or dirt. Wipe it thoroughly with HI 731318 or a lint-free cloth prior to insertion. · It is important that the sample does not contain any debris. · It is possible to take multiple readings in a row, but it is recommended that a zero reading be taken for each sample and that the same cuvet is used for zeroing and measurement. · It is important to discard the sample immediately after the reading is taken because the glass might become permanently stained. [. . . ] Alkalinity above 250 mg/L or acidity above 150 mg/L will not reliably develop the full amount of color or it may rapidly fade. To resolve this, neutralize the sample with diluted HCl or NaOH. In case of water with hardness greater than 500 mg/L CaCO3, shake the sample for approximately 1 minute after adding the powder reagent.
Battery Replacement
Battery replacement must only take place in a non-hazardous area using a 9V alkaline battery. Simply slide off the battery cover on the back of the meter. [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE HANNA INSTRUMENTS HI 93701
Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service.