If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Lastmanuals provides you a fast and easy access to the user manual HANNA INSTRUMENTS HI 96770. We hope that this HANNA INSTRUMENTS HI 96770 user guide will be useful to you.
Lastmanuals help download the user guide HANNA INSTRUMENTS HI 96770.
Manual abstract: user guide HANNA INSTRUMENTS HI 96770
Detailed instructions for use are in the User's Guide.
[. . . ] Once this indication is displayed, the meter will lock up. Change the battery and restart the meter.
Measurement procedure:
Measurement
Preliminary examination:
Please examine this product carefully. Make sure that the instrument is not damaged. If any damage occurred during shipment, please notify your Dealer. [. . . ] Battery low: The battery voltage is getting low and the battery needs to be replaced.
10
Dimensions Weight
11
www. hannainst. com
This instrument is in compliance with the Directives
13·Wait for 2 minutes and then press /TIMER. READ /TIMER The instrument status indicator will appear depending on the measurement phase. 14· The instrument directly displays concentration in mg/L of silica (SiO2) on the LCD. To convert the reading to mg/L of silicon (Si) multiply the reading by 0. 467.
13
If the value is found out of specifications, please check that the cuvets are free of fingerprints, oil or dirt and repeat validation. If results are still found out of specifications then recalibrate the instrument.
CALIBRATION
Note: It is possible to interrupt the calibration procedure at any time by pressing CAL CHECK or ON/OFF. 1· Turn the meter on by pressing ON/OFF. 2· When the beeper sounds briefly and the LCD displays dashes, the meter is ready. 3· Press and hold CAL CHECK for three seconds to enter calibration mode. The display will show "CAL" during calibration procedure. The blinking "ZERO" asks for instrument zeroing.
12· Press GLP/ to edit the desired year 12-13 (2000-2099). If the key is kept pressed, the year number is automatically increased. 13· When the correct year has been set, press ZERO/CFM or READ /TIMER to confirm. Now the display will show the month blinking. or 14· Press GLP/ to edit the desired month (01-12). If the key is kept pressed, the month number is automatically increased. 15· When the correct month has been set, press ZERO/CFM or READ /TIMER to confirm. Now the display will show the day blinking. 16· Press GLP/ to edit the desired day (01-31). If the key is kept pressed, the day number is automatically increased. Note: It is possible to change the editing from day to year and to month by pressing READ /TIMER. 17· Press ZERO/CFM to save the calibration date. 18· The instrument displays "Stor" for one second and the calibration is saved. [. . . ] 6· Press ZERO/CFM and the instrument status indicator will appear on the display, 6-7 depending on the measurement phase. 7· After a few seconds the display will show "-0. 0-". The meter is now zeroed and ready for calibration. The blinking "READ" asks for reading calibration standard. [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE HANNA INSTRUMENTS HI 96770
Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service.