If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Lastmanuals provides you a fast and easy access to the user manual HANNA INSTRUMENTS HI 981410. We hope that this HANNA INSTRUMENTS HI 981410 user guide will be useful to you.
Lastmanuals help download the user guide HANNA INSTRUMENTS HI 981410.
Manual abstract: user guide HANNA INSTRUMENTS HI 981410
Detailed instructions for use are in the User's Guide.
[. . . ] If you need additional technical information, do not hesitate to e-mail us at tech@hannainst. com. These instruments are in compliance with the
FUNCTIONAL DESCRIPTION
TDS Range Resolution
SPECIFICATIONS
HI 981409 0 to 1999 mg/L (ppm) 1 mg/L (ppm) ±2% f. s. ±2% f. s.
Accuracy (@20°C/68°F) Typical EMC Deviation
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it carefully. If any damage has occurred during shipment, immediately notify your dealer or the nearest Hanna Customer Service Center. [. . . ] They also float so that they can be quickly retrieved if accidentally dropped in the tank. The supplied HI 1220 probe incorporates a fast responding temperature sensor and is made with a rugged body ideal for measurements in fertilizer mix. It measures 24 cm (9. 4") and can even be used to mix the solution!The DIN connector is protected against humidity and spills by a rubber sheath. · Ideal for hydroponics and greenhouses · Waterproof housing floats for quick retrieval · Automatic calibration · Auto shut-off for longer battery life · Fertilizer mixer (long and rugged probe) 1. ON/OFF button, to switch the meter ON and OFF, and to enter the calibration mode 4. HI 1220 replaceable conductivity probe 7. Air-vent hole Range Resolution
EC
HI 981410 0. 00 to 10. 00 mS/cm 0. 01 mS/cm ±2% f. s. ±2% f. s.
Accuracy (@20°C/68°F) Typical EMC Deviation
Temperature Compensation Automatic, 5 to 50°C (41 to 122°F), with =2%/°C Calibration Probe Automatic at 5. 00 mS/cm HI 1220 replaceable EC/TDS probe, 240 mm (9. 4'') long, with DIN connector (included)
Battery Type / Life 2 x 1. 5V AA / approx. 300 hours of use; auto-off after 5 minutes of non-use Dimensions (meter only) 86 x 94 x 33 mm (3. 4 x 3. 7 x 1. 3") Weight 150 g (5. 3 oz. )
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner.
Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
OPERATIONAL GUIDE
PROBE CONNECTION & MAINTENANCE
In order to protect the DIN connector against vapors and humidity, the meter is provided with a waterproof sheath. · Slide the protective sheath down, connect the HI 1220 probe to the DIN connector and then slide the protective sheath back up. To ensure maximum protection, make sure the connector is completely covered.
TDS TDS
CALIBRATION
For better accuracy, a frequent calibration is recommended. Moreover, the instrument should be recalibrated at least once a month, after cleaning the probe or replacing the batteries. · Press and hold the ON/OFF button for approx. The display will start blinking EC 5. 00 "1500 ppm" for HI 981409 and "5. 00 ms" for HI 981410 to indicate that the calibration mode has been entered. · Immerse the probe in the proper calibration solution: use HI 70442 (1500 TDS ppm) for HI 981409, or HI 70039 (5. 00 mS/cm) for HI 981410. · Stir gently and wait for approximately 20 seconds for the meter to perform the automatic calibration.
ON/OFF
BATTERY REPLACEMENT
When the batteries are weak, EC and TDS will display the warning symbol "Eb". To replace the batteries, remove the rear panel and replace both batteries, while paying attention to their polarity. Batteries should only be replaced in a safe area and using 1. 5V AA alkaline batteries.
CE DECLARATION OF CONFORMITY
DECLARATION OF CONFORMITY
We Hanna Instruments Srl Via E. Fermi, 10 35030 Sarmeola di Rubano (PD) ITALY herewith certify that the EC / TDS meter
HI 981409
IEC 801-2 IEC 801-3 IEC 801-4 EN 55022 EN 61010-1
HI 981410
has been tested and found to be in compliance with the following regulations: Electrostatic Discharge RF Radiated Fast Transient Radiated, Class B Electrical Safety
800
ON/OFF
800
ON/OFF
1500
ppm
ON/OFF
Date of Issue: 26-10-1999 D. Volpato - Engineering Manager On behalf of Hanna Instruments S. r. l.
· If the meter is not used for a long time, it is recommended to clean the internal sensitive pins with alcohol. [. . . ] Avoid touching this metal band at all times. During operation, ESD wrist straps should be worn to avoid possible damage to the probe by electrostatic discharges. Any variation introduced by the user to the supplied equipment may degrade the instrument's EMC performance. To avoid electrical shock, do not use this instrument when voltage at the measurement surface exceed 24 Vac or 60 Vdc. [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE HANNA INSTRUMENTS HI 981410
Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service.