Detailed instructions for use are in the User's Guide.
[. . . ] Instruction Manual
pH 300 pH 301 pH 302
Microprocessor GLP Bench-top pH/mV/ISE/°C Meters
Rel mV ppm pH
DATE TIME
CFM MEM WRONG
°C mV ppm pH INTV
CAL LOG
1
BUF 2
3
m s
ON
OFF
PAPER
PRINT
pH302
pH METER
RANGE
CAL
CFM /
TIME
PGM
ENTER /
http://www. hannainst. com
These Instruments are in Compliance with the CE Directives
1
Dear Customer, Thank you for choosing a Hanna Instruments Product. Please read this instruction manual carefully before using the instrument. This manual will provide you with all the necessary information for the correct use of the instrument, as well as a precise idea of its versatility in a wide range of applications. These instruments are in compliance with CSA, UL and (EN 50081-1 and EN 50082-1) directives.
PRELIMINARY EXAMINATION Remove the instrument from the packing material and examine it to make sure that no damage has occurred during shipping. [. . . ] Press CFM.
· Complete your temperature calibration by pressing CAL if you do not want to perform the mV calibration (see page 44).
CAL
PROCEDURE FOR pH 300 · Switch the pH 300 on while pressing CAL. The "CAL" indicator will be lit. The secondary LCD section will show "0. 0°C".
ON
CAL
PROCEDURE FOR pH 301 · Switch the pH 301 on while pressing CFM. The "CAL" indicator will be lit. The secondary LCD section will show "0. 0°C".
ON
CFM
°C
CAL
°C
CAL
· Immerse the temperature probe in the vessel with the ice and water.
0 °C (32 °F)
°C
· Immerse the temperature probe in the vessel with the ice and water and wait for a few minutes.
°C
0 °C (32 °F)
52
53
· The secondary LCD will display "0. 0 °C" with the "CAL" indicator. · As soon as the reading is stable the "CFM" indicator will start blinking.
°C
CAL
PROCEDURE FOR pH 302 · Switch the pH302 on while pressing CAL. The "CAL" indicator will be lit. The secondary LCD section will show "0. 0°C".
ON
CAL
°C
CFM
CAL
· Press the CFM key to confirm the calibration.
CFM
· Immerse the temperature probe in the vessel with hot water and wait for a few minutes. · The secondary LCD will display "5 0 . 0 °C" with the "CAL" indicator. · As soon as the reading is stable the "CFM" indicator will start blinking.
°C
50 °C (122 °F)
· Immerse the temperature probe in the vessel with the ice and water and wait for a few minutes. · As soon as the reading is stable the "CFM" indicator will start blinking. · Press CFM to confirm the calibration.
°C
0 °C (32 °F)
CFM
CFM/
°C
CAL
CFM
· Immerse the temperature probe in the vessel with hot water and wait for a few minutes.
CFM
°C
50 °C (122 °F)
· Press CFM to confirm the calibration. · The temperature calibration is complete and the meter will return to the normal operating mode.
54
· The secondary LCD will display "50. 0 °C" with the "CAL" indicator. · As soon as the reading is stable the "CFM" indicator will start blinking.
55
°C
CAL
CFM
CFM/
· Press CFM to confirm the calibration.
LOGGING WITH pH 301 This function permits the automatic logging of pH or mV or ION measurements together with the temperature without needing an operator and for a long period of time. The lot number goes from 1 to 99 and then back to #1. The maximum capacity per lot is 8000 samples. Set the appropriate logging interval (see page 14) selectable between 1, 15, 30 seconds or 1, 5, 30, 60, 120, 180 minutes. Select the appropriate measurement mode by pressing the pH or mV or ION key.
pH mV ION
· The temperature calibration is complete and the meter will return to the normal operating mode.
Press LOG and then CAL to enter the logging mode.
LOG
CAL
Once in the logging mode, the interval cannot be changed. LOG CAL Exit the logging mode first (by pressing LOG and then CAL) before setting a new interval. During logging, you can check some information about the logged data.
56
57
Press LOG and the primary LCD will show the current lot number and the secondary LCD will display the current page number.
LOG
LOG
Press LOG again to display the current sample number (the number of values that have been taken in the current lot).
LOG
when this happens the LCD will show page "0", indicating the overwriting has happened. Each time a new logging period starts, it automatically starts a new page. [. . . ] Use: general purpose.
12 mm 0. 5"
HI 4430 120 mm 4. 7"
HI 5311
120 mm 4. 7"
"S" VERSION
Kynar
®
is registered Trademark of "Pennwalt Corp. "
Ultem
®
is registered Trademark of "General Electrics Co. "
106
107
EXTENSION CABLES FOR SCREWTYPE ELECTRODES ONLY (SCREW TO BNC CONNECTOR) HI7855/1 Extension cable 1m (3. 3') long HI7855/3 Extension cable 3m (9. 9') long HI7855/5 Extension cable 5m (16. 5') long HI7855/10 Extension cable 10m (33') long HI7855/15 Extension cable 15m (49. 5') long
HI 7855 SERIES CABLE CONNECTORS CONNECTOR AND 3. 0 mm (0. 12") CABLE WITH BNC
ELECTRODE APPLICATION REFERENCE GUIDE
Application
1. Fruit
Electrodes
HI 1332B, HI 1911B, HI1312S HI 1910B, HI 1130B, HI1110S HI 1131B, HI 1111S HI 2031B, FC200B, HI 2020S, FC200S FC200B, FC200S FC911B, FC100B HI 1910B, HI 1912B HI 1053B, HI1050S HI 1230B, HI1210S HI 1331B, HI1310S FC911B, FC100B FC 200B, FC220B, FC 200S FC210B HI 1130B, HI 1912B, HI1110S HI 1135B, HI 1115S HI 1053B, HI1050S HI 1053B, FC200B, HI1050S, FC200S HI 1134B, HI 1135B, HI 2114B, HI 2910B, HI1114S, HI1115S HI 1131B, HI 1230B, HI 1332B, HI 1330B HI1111S, HI 1210S, HI 1312S, HI1310S HI1413B, HI1410S FC100B FC200B, HI 2031B, FC200S, HI2020S HI 1083B, HI1080S FC210B HI 1053B, HI 1050S HI 1413B, HI 1410S HI 1230B, HI 1210S HI 1332B, HI1312S FC200B, HI 2031B, FC200S, HI2020S HI 2031B, HI 2020S HI 1413B, HI 1410S HI 1230B, HI1210S HI 1043B, HI1040S HI 1043B, HI1040S HI 1130B, HI1110S HI 1413B, HI1410S H 1130B, HI 2114B, HI 2910B HI 1131B, HI 1111S HI 1131B, HI 1111S FC911B, HI 1912B, HI 1911B HI 1330B, HI1310S FC 220B
CONNECT TO SCREW TYPE ELECTRODES
CONNECT TO THE BNC SOCKET OF THE METER
POWER UNITS HI 710005 115VAC-12VDC voltage adapter HI 710006 230VAC-12VDC voltage adapter
13. Micro plate sampling of less than 100 mL 24. Milk and Yogurt
OTHER ACCESSORIES ChecktempC Thermometer with penetration probe (range -50. 0 to 150. 0°C) HI 76405 Electrode holder HI 7669/2W Temperature probe with 1 m (3. 3') screened cable HI 710032 Pack of 10 plain paper spare rolls (for pH 302 only) HI 710033 Replacement ink cartridge (for pH 302 only) HI 92000 Windows® compatible application software HI 920010 9 to 25-pin PC connection cable HI 920010/9 9 to 9-pin PC connection cable MANpH300R1 Instruction manual
Windows® is registered Trademark of "Microsoft Co. "
25. [. . . ]