Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... DON'T FORGET : ALWAYS READ THE USER GUIDE BEFORE BUYING !!!
If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Lastmanuals provides you a fast and easy access to the user manual KEITHLEY 2000. We hope that this KEITHLEY 2000 user guide will be useful to you.
Lastmanuals help download the user guide KEITHLEY 2000.
Detailed instructions for use are in the User's Guide.
[. . . ] only) w w w. keithley. com
A
G R E A T E R
M E A S U R E
O F
C O N F I D E N C E
General-purpose instrument that’s as easy to operate as it is to afford
2000
6½-Digit Multimeter
SCANNER OPTION 2000-SCAN
HI LO
General-purpose instrument that’s as easy to operate as it is to afford
Ch. 5 HI LO
*
HI LO HI LO
Out A
Out B
4-pole Ch. 6 HI LO
2-pole
Optional Multiplexer Cards Creating a self-contained multipoint measurement solution is as simple as plugging a scanner card into the option slot on the 2000’s back panel. This approach eliminates the complexities of triggering, timing, and processing issues and helps reduce test time significantly. [. . . ] Maximum Voltage Between Any Terminal and Model 2001 Input LO: 200V peak. ENVIRONMENTAL: Meets all Model 2000 environmental specifications. DIMENSIONS, WEIGHT: 21mm high × 72mm wide × 221mm deep (0. 83 in. 1
2000
DC Characteristics
Conditions: MED (1 PLC)1 or SLOW (10 PLC) or MED (1 PLC) with filter of 10
6½-Digit Multimeter
Function Voltage
Range 100. 0000 1. 000000 10. 00000 100. 0000 1000. 000 100. 0000 1. 000000 10. 00000 100. 0000 1. 000000 10. 00000 100. 0000 10. 00000 100. 0000 1. 000000 3. 00000 1 3. 00000 10. 00000 10. 00000 mV V V V V9 Ω kΩ kΩ kΩ MΩ16 MΩ11, 16 MΩ11, 16 mA mA A A kΩ V V V
Resolution 0. 1 µV 1. 0 µV 10 µV 100 µV 1 mV 100 µΩ 1 mΩ 10 mΩ 100 mΩ 1 Ω 10 Ω 100 Ω 10 nA 100 nA 1 µA 10 µA 100 mΩ 10 µV 10 µV 10 µV
Test Current or Burden Voltage (±5%)
Accuracy: ±(ppm of reading + ppm of range) (ppm = parts per million) (e. g. , 10ppm = 0. 001%) Input Resistance > 10 GΩ > 10 GΩ > 10 GΩ 10 MΩ ±1% 10 MΩ ±1% 24 Hour 14 23°C ± 1° 30 + 30 15 + 6 15 + 4 15 + 6 20 + 6 30 + 30 20 + 6 20 + 6 20 + 6 20 + 6 150 + 6 800 + 30 60 + 30 100 + 300 200 + 30 1000 + 15 40 + 100 20 + 6 20 + 6 20 + 6 90 Day 23°C ± 5° 40 + 35 25 + 7 20 + 5 30 + 6 35 + 6 80 + 40 80 + 10 80 + 10 80 + 10 80 + 10 200 + 10 1500 + 30 300 + 80 300 + 800 500 + 80 1200 + 40 100 + 100 30 + 7 30 + 7 30 + 7 1 Year 23°C ± 5° 50 + 35 30 + 7 30 + 5 45 + 6 45 + 6 100 + 40 100 + 10 100 + 10 100 + 10 100 + 10 400 + 10 1500 + 30 500 + 80 500 + 800 800 + 80 1200 + 40 120 + 100 40 + 7 40 + 7 40 + 7
Temperature Coefficient 0°–18°C and 28°–50°C 2+6 2+1 2+1 5+1 5+1 8+6 8+1 8+1 8+1 8+1 95 + 1 900 + 1 50 + 5 50 + 50 50 + 5 50 + 5 8+1 8+1 8+1 8+1
Model 2000 specifications
Resistance 15
Current
Continuity 2W Diode Test
1 mA 1 mA 100 µA 10 µA 10 µA 700 nA // 10MΩ 700 nA // 10MΩ < 0. 15 V < 0. 03 V < 0. 3 V <1 V 1 mA 1 mA 100 µA 10 µA
DC OPERATING CHARACTERISTICS 2
Function DCV (all ranges), DCI (all ranges), and Ohms (<10M range) Digits 6½ 3, 4 6½ 3, 7 6½ 3, 5 5½ 3, 5 5½ 5 5½ 5 4½ 5 Readings/s 5 30 50 270 500 1000 2000 PLCs 8 10 1 1 0. 1 0. 1 0. 04 0. 01
SPEED AND NOISE REJECTION
Rate 10 PLC 1 PLC 0. 1 PLC 0. 01 PLC DC NOTES
1. Add the following to “ppm of range” uncertainty:1V and 100V, 2ppm; 100mV, 15ppm; 100Ω, 15ppm; 1kΩ– <1MΩ, 2ppm; 10mA and 1A, 10ppm; 100mA, 40ppm. Speeds are for 60Hz operation using factory default operating conditions (*RST). Autorange off, Display off, Trigger delay = 0. Speeds include measurement and binary data transfer out the GPIB. 1 PLC = 16. 67ms @ 60Hz, 20ms @ 50Hz/400Hz. The frequency is automatically determined at power up. For signal levels >500V, add 0. 02ppm/V uncertainty for the portion exceeding 500V. Must have 10% matching of lead resistance in Input HI and LO. For 2-wire ohms, add 1Ω additional uncertainty. For rear inputs, add the following to temperature coefficient “ppm of reading” uncertainty 10MΩ 95ppm, 100MΩ 900ppm. Operating environment specified for 0º to 50ºC and 50% RH at 35ºC.
Readings/s 5 50 500 2000
Digits 6½ 6½ 5½ 4½
RMS Noise 10V Range < 1. 5 µV < 4 µV < 22 µV < 150 µV
NMRR 12 60 dB 60 dB — —
CMRR 13 140 dB 140 dB 80 dB 80 dB
DC SYSTEM SPEEDS 2, 6
RANGE CHANGE 3: 50/s. ASCII READINGS TO RS-232 (19. 2K BAUD): 55/s. EXTERNAL TRIGGER RATE: 400/s.
DIGITAL MULTIMETERS & SYSTEMS
DC GENERAL
LINEARITY OF 10VDC RANGE: ±(1ppm of reading + 2ppm of range). DCV, Ω, TEMPERATURE, CONTINUITY, DIODE TEST INPUT PROTECTION: 1000V, all ranges. MAXIMUM 4WΩ LEAD RESISTANCE: 10% of range per lead for 100Ω and 1kΩ ranges; 1kΩ per lead for all other ranges. DC CURRENT INPUT PROTECTION: 3A, 250V fuse. SHUNT RESISTOR: 0. 1Ω for 3A, 1A, and 100mA ranges. CONTINUITY THRESHOLD: Adjustable 1Ω to 1000Ω. [. . . ] Applies to 0°–18°C and 28°–50°C. For signal levels >2, 2A, add additional 0. 4% to “of reading” uncertainty. Bench Configuration (with handle and feet): 104mm high × 238mm wide × 370mm deep (4. 13 in × 9. 38 in × 14. 56 in). SHIPPING WEIGHT: 5kg (11 lbs).
Remote Interface
GPIB (IEEE-488. 1, IEEE-488. 2) and RS-232C.
Frequency and Period Characteristics 1, 2
ACV Range 100 mV to 750 V Frequency Range 3 Hz to 500 kHz Period Range 333 ms to 2 µs Gate Time 1 s (SLOW) Resolution ±(ppm of reading) 0. 3 Accuracy 90 Day/1 Year ±(% of reading) 0. 01
FREQUENCY NOTES
1. [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE KEITHLEY 2000
Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service.
Click on "Download the user manual" at the end of this Contract if you accept its terms, the downloading of the manual KEITHLEY 2000 will begin.