[. . . ] SPECTRORADIOMETER CS-2000
Instruction Manual
Safety Symbols
The following symbols are used in this manual to prevent accidents which may occur as a result of incorrect use of the instrument.
Denotes a sentence regarding a safety warning or note. Read the sentence carefully to ensure safe and correct use. The operation must never been performed. The instruction must be strictly adhered to. [. . . ] The setting of absolute value (ABS) and difference (DIFF) is saved even after switching OFF (O).
41
MEAS
UC00 T01
cd 2 m
Lv x y Lv %Lv x. 0001
cd 2 m
Obs SPD SYNC L E N ND BL 2° Nr m N o n S t d N o n o n
Setting
MEAS
SNGL
UC00 T01
When Using Closeup Lens
Use the closeup lens in the optional accessories to measure a fine surface. For the attachment of the closeup lens, refer to the instruction manual for the closeup lens. If the closeup lens is attached, the measurement value must be compensated for the lens transmittance. This compensation coefficient is attached to the closeup lens. Use the data management software CS-S10w Professional supplied with the CS-2000 as standard to set the coefficient in this instrument. After that, when the closeup lens is selected as lens type, the measurement value compensated with the compensation coefficient can be obtained. See instruction manual of CS-S10w for details. The wrong setting of lens type leads to inaccurate measurement. Do not use the closeup lens with the ND filter. It leads to inaccurate measurement.
* Lens type : STANDARD, CLOSE-UP * Factory default setting : STANDARD
Setting
Operation Procedure
6, 7
1 2, 3, 4, 5
1.
Press MENU key when the MEAS screen is displayed.
The MENU screen appears. When the backlight of the LCD has been turned off via BACKLIGHT key on the MEAS screen, the backlight is turned on.
MEAS
SNGL
UC00 T01 cd m2
Lv x y
35. 4060
MENU
Obs SPD SYNC L E N ND BL 2° Nr m I n t S t d N o n o n
MEAS MEMORY TARGET OPTION SETUP
2.
Press either or key to select [OPTION] and then press ENTER key.
The MENU OPTION screen appears. The current set contents are displayed in the [LENS] item.
42
MENU OPTION USER. CAL LENS ND-F I L T E R NON STD NON
3.
Press either or key to select [LENS] and then press ENTER key.
The MENU - SETUP LENS (Lens type selection) screen appears.
MENU OPTION LENS STANDARD CLOSE- UP
4.
Press either [CLOSE-UP].
or
key to select
MENU OPTION LENS STANDARD CLOSE- UP
When you removed the closeup lens, select [STANDARD].
5.
Press ENTER key.
When the lens type is set, the MENU - OPTION screen appears again on the LCD. By pressing ESC key, after the setting is canceled the MENU - OPTION screen appears again on the LCD. Lens type setting is saved even after switching OFF (O).
LENS
MENU OPTION USER. CAL NON C- UP NON
ND-F I L T E R
6.
Press ESC key.
The MENU screen appears again on the LCD.
7.
Press ESC key.
The MEAS screen appears again on the LCD.
MEAS
SNGL
UC00 T01 cd m2
Lv x y
35. 4060
Obs SPD SYNC L E N ND BL 2° Nr m I n t C- U N o n o n
43
Setting
When Using ND Filter
Use the ND filter in the optional accessories when measuring a high-luminance object. If the ND filter is attached, the measurement value must be compensated for the filter transmittance. This compensation coefficient is attached to the ND filter. Use the data management software CS-S10w Professional supplied with the CS-2000 as standard to set the coefficient in this instrument. [. . . ] not finish in the set measurement time.
10
83
Explanation
Has the closeup lens been attached?
Error Symptom No. 11 Actual measurement time is different from the displayed measurement time.
Item to Check
Corrective Action
Reference Page
12
Measurement value on the LCD disappears.
13
During USB communication: Cannot download data output from this instrument on the PC. Cannot enter commands or data from Has remote mode been the PC to this canceled?instrument.
The displayed measurement time is the remaining time. [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE KONICA CS-2000
Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service.