If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Lastmanuals provides you a fast and easy access to the user manual ROTOTHERM CLEARSCAN RECORDER. We hope that this ROTOTHERM CLEARSCAN RECORDER user guide will be useful to you.
Lastmanuals help download the user guide ROTOTHERM CLEARSCAN RECORDER.
Manual abstract: user guide ROTOTHERM CLEARSCAN RECORDER
Detailed instructions for use are in the User's Guide.
[. . . ] user manual
installation, setup, operation & calibration
With over sixty years of experience, British Rototherm continues to be a world leader in the manufacture of industrial instrumentation for measuring, monitoring and controlling Temperature, Pressure, Humidity, Flow and Level. British Rototherm products are manufactured in its modern factory located in the South Wales region of the United Kingdom. Sited on 6 acres of land at Kenfig Industrial Estate, the company operates from a modern 6, 030 square metre (65, 000 square feet) factory. Suitable for many industrial applications, Rototherm products are precision built to the highest quality. [. . . ] Either type is terminated in a tapped hole to which is attached the adjustable horizontal link via a shoulder screw forming a pivot without excess side movement for smooth friction free movement. The adjustable horizontal link is screwed into the ratio post at its opposite end providing a smooth pivot without excessive free movement or play thereby eliminating hysteresis. The ratio post provides an adjustable point of pivot for the adjustable horizontal link and span adjustment is thereby provided by the position of the ratio post which is fitted through the slot in the pen bracket and clamped by a ratio post washer and locknut in terms of its pivot distance to the pen bracket spindle. To replace a pen mechanism assembly, remove the shoulder screw(s) from the ratio post(s) and undo the two nuts securing the pen mechanism bracket to the mounting bracket. Retain all fixings ensuring the relative shoulder screw is mated with its original linkage. Mount a new mechanism in reverse order and measure the ratio post(s) distance from the spindle. Adjust the ratio post(s) to this dimension before carrying out the appropriate calibration procedure.
12
BASIC FAULT FINDING
Recorder pen inaccurate or gives no indication Probable cause Action Probable cause Action Probable cause Action Measuring element broken, pressure line plugged or broken. Check element / pressure line; replace system if necessary. Repair and re-connect if possible; replace if necessary. Recorder out of calibration; measuring element damaged. Set-up and recalibrate; replace system if necessary.
No record or poor trace on chart Probable cause Action Probable cause Action Chart not rotating Probable cause Action Probable cause Action Clock unwound. Adjust pen friction.
After carrying out the basic fault finding if the problem is still evident, refer to the detailed maintenance procedures or contact British Rototherm or your local authorised British Rototherm distributor for spares, information and technical advice and support.
13
ClearscanTM Recorder Series - Operating and Service Manual
TEMPERATURE BOURDON
COARSE ZERO ADJUSTMENT SCREW
COMPENSATOR ARM
RATIO POST (UNDERNEATH PEN ARM)
LINEARITY LINK
CATCH POST
PENLIFTER
CHART DRIVE CHART PLATE (PARTIALLY REMOVED)
CAPILLARY
OUTLET PLATE
CAPILLARY TUBING
SENSING BULB
Fig. 5a Temperature Recorder Internal Layout
14
COARSE ZERO ADJUST SCREW BOURDON TUBE TAKE OFF ASSEMBLY
ZERO ADJUST SCREW
LINEARITY LINK
CATCH POST
PENLIFTER
CHART DRIVE
CHART PLATE (PARTIALLY REMOVED)
CAPILLARY
OUTLET PLATE
PRESSURE CONNECTION
Fig. 5b Pressure Recorder Internal Layout (ranges up to 40 bar)
15
ClearscanTM Recorder Series - Operating and Service Manual
ADJUSTABLE HORIZONTAL LINK
DRIVE ARM
CATCH POST COIL LEVER SCREW COARSE ZERO ADJUST BOURDON TUBE
PENLIFTER
CHART DRIVE
CHART PLATE (PARTIALLY REMOVED)
CAPILLARY
OUTLET PLATE
PRESSURE CONNECTION
Fig. 5c Pressure Recorder Internal Layout (ranges 60 bar and above)
16
RANGE ADJUST RATIO POST BENEATH PEN ARM
HORIZONTAL ADJUSTABLE LINK
DRIVE ARM
DRIVE ARM ADJUST SCREW
PEN STOP
TORQUE TUBE CLAMP SCREW
FLANGED BEARING
CALIBRATION CHART SHAPE BALANCE WEIGHT
0% INDICATION
CHART TIME LINE
50% INDICATION 100% INDICATION
CHART DRIVE HUB SHOWN LOCKED
Fig. 5d Differential Pressure Recorder Internal Layout
17
ClearscanTM Recorder Series - Operating and Service Manual
TEMPERATURE PEN CALIBRATION
Mechanical calibration is concerned with the accurate setting of zero, span and linearity of the recording pen in comparison with a known temperature read with a certified thermometer traceable to national calibration standards. The adjustments detailed below must be performed in the order listed and the instrument must be allowed to settle before readings and adjustments are made. Unless a replacement mechanism assembly, temperature system or mechanical linkage is to be fitted it should not be necessary to make coarse or linearity adjustments. If it is not possible to achieve one or more of the following settings, contact your nearest Rototherm Distributor or the company itself. Allow the recorder to stabilise with the measuring element in a low temperature just above scale minimum. The temperature recording pen should read within ±1% of full span at the correct temperature value. If necessary, adjust pen fine zero adjust screw to bring the pen to the correct reading. [. . . ] If the pen deviates less than 10% from zero in either direction, perform fine adjustment by using the pen zero adjust screw and then proceed to step 11b. If the pen deviates by more than 10% from zero, make adjustments as follows in step 11a: Slacken the centre screw of the torque tube shaft whilst supporting the drive arm, slip the pen until it indicates zero and re-tighten. Apply 100% pressure and observe the pen. Make the necessary adjustment by slackening slightly the range adjust locknut enough to slide the ratio post. [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE ROTOTHERM CLEARSCAN RECORDER
Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service.