Reverse osmosis equipment for dish/glass washmachines.
use and maintenance
SERIAL N°: __________________ CODE: Note: __________________ _____________________________________________
MANUAL
-1-
INDEX
1 1. 1 1. 2 1. 3 1. 4 1. 5 2 2. 1 2. 2 2. 3 2. 4 2. 5 3 4 4. 1 4. 2 4. 3 4. 4 5 5. 1 5. 2 5. 3 5. 4 5. 5 5. 6 5. 7 5. 8 6 6. 1 6. 2 6. 3 6. 4 7 GENERAL INFORMATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 MANUFACTURER / ASSISTANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 CERTIFICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 AIM AND CONTENTS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . [. . . ] Before opening the machine, remove covers and/or panels and/or protection grids and make sure the switch is off and the machine is unplugged. ATTENTION When cleaning the machine do not use corrosive products, acids, abrasive or steel brushes. Do not directly wash the machine with high-pressure jets of water. 5. 3 REPLACING FILTER CARTRIDGE
ITEM TO CHECK
The repleace the prefilter follow these steps: Remove the power plug. Remove the filter cartridge, turning it left ¼. Install new cartridge in the machine; take caref about the filter o-rings. Reset the electronic board (if used with time limit funciont) (see § 4. 4. 3). Sterilize the system (see § 5. 8).
5. 4
PRELOAD EXPANSION TANK
To check the preload of the pressure tank, follow these steps: Remove the power plug. Empty the expansion tank delivering water to the permeate auxiliary outlet. Remove the side panels and identify the connection of the preload in the tank. With a compressor and a pneumatic gun, bring the preload pressure at 2. 5bar Install the panels and connect the power plug.
- 14 -
5. 5
VERIFICATION OF THE SETTING
For proper operation of the machine, check periodically the calibrations. All the calibrations must be performed by qualified personnel. 5. 5. 1 Setting pressure switch Install a pressure gauge (0-6 bar or similar) between the permeate outlet and the faucet. Screw two turns the calibration nut of the pressure switch. Open the knob of the tap, after which it slowly close the tap until the pressure gauge reaches about 3. 5 bar. Slowly unscrew the calibration nut of pressure switch until the operation of the machine becomes discontinuous. Apply the glue sealant to calibration nut.
v re o ite g lazio e n p s o tato re s s
5. 5. 2
Check operating pressure Install a T fitting and a pressure gauge 0-10 bar (or similar) on the pump outlet line. The operating pressure must be between 6 and 7 bar.
5. 5. 3
Calibrating conductivity probe Open the electronic board box. After 2 minutes, take a sample and measuring the conductivity with a reference instrument properly calibrated. Turn the conductivity calibration potentiometer until the displayed value matches the value measured by the instrument.
5. 6
CLEANING CONDUCTIVITY PROBE CONNECTORS Disconnect the connectors of the conductivity probe. Clean the male and female connectors from possible oxidation. Tighten the connector with pliers and reassembly.
- 15 -
5. 7
CLEANING SOLENOID VALVES FILTER Close the inlet water valve. [. . . ] O símbolo da barra no barril lembra ao utilizador de não jogar este producto como um lixo municipal não classificado. (D) Gemäß Richtlinie 2002/96/EC wird diese Produkt als elektrisches/elektronisches Gerät eingestuft. Dieses Gerät nicht als nicht klassifizierten Stadtmüll beseitigen. Diese Gerät an eine dafür vorgesehene Sammelstelle bringen, gemäß den vorortigen und auf nationalem Gebiet geltenden Gesetzesbestimmungen. [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE SMEG WO-04
Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service.