[. . . ] USER'S GUIDE
Vaisala DRYCAP® Dewpoint Transmitter DMT142
M210397EN-D
PUBLISHED BY Vaisala Oyj P. O. Box 26 FIN-00421 Helsinki Finland Phone (int. ): +358 9 8949 1 Fax: +358 9 8949 2227
Visit our Internet pages at http://www. vaisala. com/ © Vaisala 2006 No part of this manual may be reproduced in any form or by any means, electronic or mechanical (including photocopying), nor may its contents be communicated to a third party without prior written permission of the copyright holder. The contents are subject to change without prior notice.
Please observe that this manual does not create any legally binding obligations for Vaisala towards the customer or end user. All legally binding commitments and agreements are included exclusively in the applicable supply contract or Conditions of Sale.
_________________________________________________________________________________
Table of Contents
CHAPTER 1
GENERAL INFORMATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . [. . . ] The possible dry end drift is then corrected to correspond to the calibrated values. During the auto-calibration the transmitter outputs the Td value prior to the procedure.
VAISALA_________________________________________________________________________ 5
User's Guide _______________________________________________________________________
Auto-calibration is carried out only if several criteria for the measurement environment are fulfilled. This ensures the reliability of the adjustments, and maintains the excellent long term stability that the patented technology offers. These criteria include for example a stable enough moisture level in the measured atmosphere. If the conditions are not fulfilled, the auto-calibration function is postponed until satisfactory conditions are reached.
Sensor Purge
Sensor purge is also an automatic procedure that minimizes the drift at the wet end readings of the dewpoint measurement. Sensor purge is performed once a day or when the power is switched on. The sensor is heated for several minutes which will then evaporate all excess molecules out of the sensor polymer. This, together with the autocalibration results in a very small drift of the sensor due to the very linear behaviour of the polymer technology.
Sensor Warming in High Humidities
Additionally the DMT142 transmitter has a warming feature which prevents the sensor and filter from becoming wet in high humidities. High humidity may be present when the dewpoint temperature rises close to the gas temperature. Sensor warming is switched on automatically when the humidity in the measured gas increases to a level where dew can start to form. The advantage of sensor warming is the rapid response of dewpoint measurement. A wet sensor and filter would otherwise result in a dewpoint equal to ambient temperature (that is RH = 100 %). If in spite of sensor warming the sensor gets soaked, it will recover fully back to normal operation after it dries out.
6 ____________________________________________________________________ M210397EN-D
Chapter 4 ________________________________________________________________ Installation
CHAPTER 4
INSTALLATION
This chapter provides you with information that is intended to help you install Dewpoint Transmitter DMT142.
Selecting Location
In the mounting of Dewpoint Transmitter DMT142 it is important that the point of installation represents well the gas to be measured. Temperature changes do not affect the dewpoint measurement, but pressure changes will have an effect on the measurement. All leaks in the system must be eliminated to avoid ambient humidity affecting the measurement.
VAISALA_________________________________________________________________________ 7
User's Guide _______________________________________________________________________
0507-094
Figure 2
Mounting Dewpoint Transmitter DMT142
The numbers in the following list refer to Figure 2 above: 1. Select a proper location to mount the DMT142 and remove the yellow sensor protective cap. Mount the transmitter to the measurement point. The transmitter has parallel thread G1/2" ISO 228/1. Connect the wires of the connection cable. When using the cable provided with DMT142 refer to the following wiring section. See the power supply requirements on page 11. Plug in the cable to the transmitter and turn on the power supply. [. . . ] If there is need for calibration or adjustment contact Vaisala Service Center or your local Vaisala representative.
Filter Change
The DRYCAP® sensor in the DMT142 transmitter is protected with a stainless steel sintered filter. The filter is recommended to be changed whenever it looks dirty or as a part of the periodical maintenance procedure. The recommended replacement interval depends heavily on the application and can vary from less than a year up to several years.
VAISALA________________________________________________________________________ 29
User's Guide _______________________________________________________________________
A dirty or corroded filter may cause increased response times. In case the DMT142 transmitter is used in a drying system with silica gel (or other drying agent) the dust from the dryer may collect on the sensor slowing the response but also generating a micro climate around the sensor altering the dewpoint at the sensor. [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE VAISALA DMT142 Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service. Click on "Download the user manual" at the end of this Contract if you accept its terms, the downloading of the manual VAISALA DMT142 will begin.