[. . . ] USER'S GUIDE
Vaisala CARBOCAP® Hand-Held Carbon Dioxide Meter GM70
M010139EN-F
PUBLISHED BY Vaisala Oyj P. O. Box 26 FIN-00421 Helsinki Finland Phone (int. ): +358 9 8949 1 Fax: +358 9 8949 2227
Visit our Internet pages at http://www. vaisala. com/ © Vaisala 2008 No part of this manual may be reproduced in any form or by any means, electronic or mechanical (including photocopying), nor may its contents be communicated to a third party without prior written permission of the copyright holder. The contents are subject to change without prior notice.
Please observe that this manual does not create any legally binding obligations for Vaisala towards the customer or end user. All legally binding commitments and agreements are included exclusively in the applicable supply contract or Conditions of Sale.
________________________________________________________________________________
Table of Contents
CHAPTER 1
GENERAL INFORMATION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . [. . . ] Press
EXIT to return to the basic display.
7.
0505-230
Figure 10
Changing the Shortcut Keys
The following number refers to Figure 10 above: 1 =
Hold/Save-shortcut key replaced by pressure setting shortcut key P
28 __________________________________________________________________ M010139EN-F
Chapter 6 __________________________________________________________________Settings
Key Click and Backlight On Key Press
1. Open the MENU, press Select Settings, press . .
OPEN.
Select User interface, press
To turn the sound effect while pressing the buttons OFF or ON, select Key click and press ON/OFF. To turn the backlight effect while pressing the buttons OFF or ON, select Backlight on key press ON/OFF. Press
EXIT to return to the basic display.
Setting Date and Time
1. Open the MENU, press Select Settings, press . .
OPEN.
Select Date and time, press
The default date presentation format is year-month-date. To change the date, select Date and press SET. If you want to change the format, select D. M. Y (date. month. year) or M/D/Y date format (month/day/year), press
ON/OFF.
5.
The default time presentation format is 24-hour clock. To change the time, select Time and press SET. If you want to change the format, select 12-hour clock, press ON/OFF. Press
EXIT.
6.
Device Information
The basic information about the indicator and the probe is found as follows: 1. Open the MENU, press Select Settings, press .
OPEN.
VAISALA _______________________________________________________________________ 29
USER'S GUIDE____________________________________________________________________
3. 4.
Select Device information, press
SHOW.
The first display gives the information on the MI70 indicator. Press MORE to get the information on the probe. Press OK and EXIT to return to the basic display.
0505-231
Figure 11
Device Information Displays
The following numbers refer to Figure 11 above: 1 2 3 4 = = = = Software version of the MI70 indicator Serial number of the MI70 indicator Software version of the probe Location of the last adjustment (alternatives: Vaisala, GM70, or GMK220)
Reverting Factory Settings
Factory settings of the indicator only can be reverted to clear all changed settings and data memory of the indicator. Reverting factory settings of the indicator does not affect the probe calibration. Factory calibration of the probe cannot be reverted if adjustment is made (see Chapter 10 Calibrating and Adjusting the Probes on page 43). Open the MENU, press Select Settings, press .
REVERT. Press YES to confirm the OPEN.
Select Factory settings, press reverting.
The power turns off automatically. When switching on again, the factory settings are reverted. You need to set the language, date, and time again. [. . . ] Pull the probe out firmly.
Changing the Probe Filter
Spare filters for the probe can be ordered from Vaisala, see list of accessories on page 62. Remove the plastic grid by pulling it out. Insert a new filter paper and plastic grid.
Cleaning
Wipe the plastic parts of the device with a damp paper or towel. Do not let water flow down to electronics through the buttonholes.
Changing the Battery Pack
New rechargeable battery packs can be ordered from Vaisala. [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE VAISALA GM70 Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service. Click on "Download the user manual" at the end of this Contract if you accept its terms, the downloading of the manual VAISALA GM70 will begin.