Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... DON'T FORGET : ALWAYS READ THE USER GUIDE BEFORE BUYING !!!
If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Lastmanuals provides you a fast and easy access to the user manual ZEISS HBO 100. We hope that this ZEISS HBO 100 user guide will be useful to you.
Lastmanuals help download the user guide ZEISS HBO 100.
Detailed instructions for use are in the User's Guide.
[. . . ] Violations will entail an obligation to pay compensation. All rights reserved in the event of granting of patents or registration of a utility model.
Published by:
Carl Zeiss MicroImaging GmbH P. O. B. 4041, 37030 Göttingen, Germany Phone: +49 (0) 551 5060 660 Fax: +49 (0) 551 5060 464 E-mail: micro@zeiss. de www. zeiss. de
Number of this manual: 46-0063 e Date of issue: 03/01/2006
B 46-0063 e 03/06
1
Carl Zeiss
Contents
Microscope Illuminating System HBO 100
Contents
1 1. 1 1. 2 1. 3 1. 4 1. 5 1. 6 1. 7 1. 8 2 2. 1 2. 2 3 3. 1 3. 2 3. 3 3. 4 3. 5 3. 6 Notes on Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Regulations concerning Instrument Safety and Electromagnetic Compatibility (EMC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . [. . . ] 417009) and the safety gloves with artery protection (Cat. For safety reasons, a safety holder is inserted in the microscope illuminator instead of the bulb during transport. Both the HBO 100 microscope illuminator and the HBO 103 lamp must be transported in their original packaging. This packaging must be kept for storage and disposal. The transport and storage temperatures to be observed correspond to those of the microscope stands.
- When they are hot, the lamps are under high internal pressure. They may only be inserted in the microscope illuminator if a safety mask and safety gloves (see chapter 1. 3) are worn. - The HBO 103 lamps are under high pressure during operation; they must always be operated in closed lamp housings!- The light of the HBO 103 short-arc lamps contains a high amount of UV light: Never look directly into the lamp or in reflected light Wear safety glasses (sunglasses) when operating the lamps Cover inspection openings with filters - The UV radiation of the lamps produces little quantities of harmful ozone. Take care to ventilate the working room with fresh air. - Both the HBO 100 microscope illuminator and the HBO 103 lamps must be protected from impact and humidity. Always keep the ventilation slats clear and ensure that no objects enter the instrument through the ventilation slats. 423010 HBO 100 illuminator, self-adjusting Cat. 423011-9901
Lamp housing
Length Width Height Connection to stand Connection to transformer Transformer
165 mm 166 mm 205 mm dovetail Cable with special plug Transformer for HBO 100 Cat. No. 432604-9901 100-240 V AC 50-60 Hz 155 VA F1/F2 2x T 2 A/H Mercury vapor short-arc lamp OSRAM Type HBO 103 W/2 Cat. 380301-9350-000 OSRAM Type HBO 100 W/2 Type HBO 103 W/2
Line connection Line frequency Power consumption Fuse Lamp
For the technical data of the lamps please see the information provided by the manufacturer. Recommended: Average life of the lamps
200 h 300 h
B 46-0063 e 03/06
9
Carl Zeiss
Instrument Description Technical Data, Line spectrum
Microscope Illuminating System HBO 100
Fig. 3
Line spectrum of the HBO 100/103 mercury short-arc lamps
10
B 46-0063 e 03/06
Microscope Illuminating System HBO 100
Start-Up Lamp Insertion and Change
Carl Zeiss
3
3. 1
Start-Up
Lamp Insertion and Change
- Before inserting the lamp, make sure the illuminating system has been disconnected from line power. 4/1), push down its clamping lever (Fig. 4/7) and put the heat sink onto the lamp (Fig. By doing so, the electric connection is established and the heat sink clamped.
Do not exert any force to the lamp, when attaching the heat sink. · Put the top part of the illuminator onto the bottom part. · Turn both fastening screws (Fig. [. . . ] 8
Adjusting elements of HBO 100 for manual adjustment
14
B 46-0063 e 03/06
Microscope Illuminating System HBO 100
Start-Up Focusing and Centering
Carl Zeiss
3. 4
Focusing and Centering the HBO 100 Microscope Illuminator, self-adjusting (423011-9901)
Automatic adjustment of the HBO 100 requires the collector to be set to mid position. · Turn the collector control knob (Fig. After the transformer of the connected HBO 100 microscope illuminator was switched on, the five minutes warm-up phase begins. After the five minutes are over, automatic adjustment is started. [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE ZEISS HBO 100
Lastmanuals offers a socially driven service of sharing, storing and searching manuals related to use of hardware and software : user guide, owner's manual, quick start guide, technical datasheets... In any way can't Lastmanuals be held responsible if the document you are looking for is not available, incomplete, in a different language than yours, or if the model or language do not match the description. Lastmanuals, for instance, does not offer a translation service.